User manual

H61.0.3X.6C-01 User’s Manual GMH 3181- page 8 of 16
Menu
Param.
Values
Description
C_log
C_dat
Stor
‚Menu‘
or
SEt
AL.
Set Alarm: Settings Of Alarm Function
AL.
on
Alarm on, with horn-sound
no.So
Alarm on, without horn-sound
oFF
Alarm deactivated
AL.Lo
p.r.t. chart
Min alarm rail (not when AL. oFF, Sensor-Min is the lower display range
of connected sensor)
AL.Hi
p.r.t. chart
Max alarm rail (not when AL. oFF, Sensor-Max is the upper display range
of connected sensor)
SEt
LoGG
Set Logger: Configuration Of Logger Function
*
*
Func
CYCL
Cyclic: logger function ‚cyclic logger‘
*
*
*
Stor
Store: logger function ‚individual value logger‘
*
*
*
oFF
no logger function
*
*
*
CYCL
1..3600
Cycle time of cyclic logger [seconds]
*
*
Lo.Po
on/oFF
Low-power logger with very low power consumption
(only for cyclic logger and slow measuring rate)
*
*
SEt
CLOC
Set Clock: Setting Of Real Time Clock
CLOC
HH:MM
Clock: Setting of time hours:minutes
dAtE
TT.MM
Date: Setting of date day.month
YEAr
YYYY
Year: Setting of year
C_log: cyclic logger (active) C_dat: cyclic logger with data (no active logging) C_stor: individual value logger with data
Note: When using the logger function some settings in the menu may not be accessible (*).
If this settings should be changed, the logger has to be stopped before, eventually the logger
data has to be cleared. (p.r.t. chapter 8)
device type
analog output
adjustment of sensor
alarm function
dAC.0 / dAC.1
OFFS
Al.Lo
Al.Hi
GMH 3181-002
-500.0 ... 500.0 Pa
-50.0 ... 50.0 Pa
- 500 Pa ... AL.Hi
AL.Lo ... 500 Pa
GMH 3181-01
-1.00 ... 25.00 mbar
-5.00 ... 5.00 mbar
- 1 mbar ... AL.Hi
AL.Lo ... 25 mbar
GMH 3181-07
-10.0 ... 350.0 mbar
-50.0 ... 50.0 mbar
- 10 mbar ... AL.Hi
AL.Lo ... 350 mbar
GMH 3181-07B
-10.0 ... 420.0 mbar
-50.0 ... 50.0 mbar
- 10 mbar ... AL.Hi
AL.Lo ... 420 mbar
GMH 3181-07H
-1.00 ... 70.00 mbar
-5.00 ... 5.00 mbar
-100 mbar ... AL.Hi
AL.Lo ... 70 mbar
GMH 3181-13
-100 ... 2000 mbar
-500 ... 500 mbar
-100 mbar ... AL.Hi
AL.Lo ... 2000 mbar
GMH 3181-13
Option: MB -1..2 BAR
-1000 ... 2000 mbar
-500 ... 500 mbar
-1000 mbar ... AL.Hi
AL.Lo ... 2000 mbar
GMH 3181-12
0 ... 1300 mbar absolute
-500 ... 500 mbar
0 bar ... AL.Hi
AL.Lo ... 1300 mbar
7 Remarks To Special Features
7.1 Different Kinds Of Measuring: „rAtE-Slo, -P.dEt, -FASt“
Three different kinds of measuring pressure are supported. Two of them are working with high measuring
frequency of more than 1000 measurings per second. If one of them was chosen in the configuration (see
above), this will be displayed in the secondary display: „P.dEt“ or „FASt“.
7.1.1 rAtE-Slo: Standard Measuring
Measuring rate 4Hz, averaging and filter functions are active.
Application: Measuring of slowly changing or static pressures, e.g. measuring of leakproofness,
atmospheric pressure...
Highest accuracy, high noise immunity (EMI and unstable measuring signals), low power consumption.
7.1.2 rAtE-P.dEt: Peak detection
Measuring rate >1000Hz, the value is displayed unfiltered.
Application with logger function: Measuring of short pressure peaks or fast changing pressures with a
resolution of < 1 ms. The cyclic logger function records the arithmetic mean value, the highest and the
lowest peak of the refering time interval.
Attention: higher power consumption, measuring is sensitive to noise (EMI,..).