H61.0.2X.6C-01 User’s Manual GMH 3161- page 7 of 12
5 Start Operation
Connect sensor, turn on device via key.
After segment test the device displays some configuration:
If a zero point adjustment was carried out the display shows shortly „nuLL Corr“.
After that the device is ready for measuring.
6 Configuration
To change device settings, press Menu (key 4) for 2 seconds. This will call the configuration menu.
Pressing key Menu jumps between the parameters.
The parameters can be changed with (key 2) or (key 5).
Quit (key 6) finishes the configuration and returns to standard measuring operation.
Parameter
Value
Description
‚Menu‘
or
mbar, bar, …
Unit: Unit of display
oFF/on
Sea level correction: on or off
(only available at GMH 3161-12)
-2000 … 9999
Altitude: Input of altitude above sea level [m] (only if SL=on)
(only available at GMH 3161-12)
1 … 120
Auto Power Off time in minutes
oFF
Auto Power Off deactivated
01, 11 … 91
Base address of interface
OFFS
refer to list below
The offset of sensor will be displaced by this value to compensate for
deviations in the probe or in the measuring device.
oFF
Zero displacement inactive (=0.00)
SCAL
-2.000 2.000
The measuring scale of sensor will be changed by this factor [%] to
compensate deviations of temperature probe or measuring device
oFF
Scale correction factor inactive (=0.000)
Device type
adjustable offset
GMH 3161-002
-50.0 ... 50.0 Pa
GMH 3161-01
-5.00 ... 5.00 mbar
GMH 3161-07
-50.0 ... 50.0 mbar
GMH 3161-07B
-50.0 ... 50.0 mbar
GMH 3161-07H
-5.00 ... 5.00 mbar
GMH 3161-13
-500 ... 500 mbar
GMH 3161-12
-500 ... 500 mbar
1 2 3
4 5 6
ON
OFF
max
Store
CAL
min
Set
Menu Quit
MPaPa
bar
mbar
kPa
Tara
AL
LoggSLPSI
mbar
MPaPa
bar
mbar
kPa
Tara AL
LoggSLPSI
MPaPa
bar
mbar
kPa
Tara AL
LoggSLPSI
MPa
Pabar
mbar
kPa
Tara AL
LoggSL
PSI
MPaPa
bar
mbar
kPa
Tara
AL
LoggSL
PSI
mbar