Instructions
H82.0.03.6C-08 Operating Manual GMH 5155 page 16 of 22
_____________________________________________________ _____________________________________________________________________________
At measuring kind "rAtE-Slo a power saving function is selectable: Lo.Po. If activated a
measurement is only done if it is also logged (i.e. no additional measurements between the saved
values). This seriously reduces the power consumption and is therefore especially appropriate for
long-term measurements (e.g. leak test) without an external power supply.
a) Start recording:
Press “store” for 2 seconds: Start Menu, press again: automatic recording will be started.
saved data set.
If the logger memory is full, the recording stops automatically and the display shows
If Low-Power-Logger-Function Lo.Po = on the device switches itself off as soon as the memory gets filled.
b) Stop recording:
Press “store” for 2 seconds:
Select with or .
Do not stop recording
(cancel menu)
Stop recording
If you try to switch off the device while cyclic recording is active you will be asked whether the
recording should really be stopped.
The device can only be switched off if the recording is stopped.
Auto-off function is deactivated as long as cyclic recording is active.
c) Clear recordings:
Press “store” for 2 seconds: If there are data sets stored and recording is already stopped
is displayed
Select with or to call the clear-menu
Select with or .
Clear nothing (cancel menu)
Clear all recordings
Clear the last recording