b) Read manual recordings:
Saved data sets can be viewed both with PC-software GSOFT3050 and directly on the device display.
Do not change the pressure sensor if manual recording is selected and values are already
stored, because otherwise invalid data may be read-out. The pressure sensor used during the
recording has to be connected!
Press “menu” for 2 seconds: is displayed

configuration menu is displayed:
Press shortly: Change between measuring values, measuring point and
date+time of the data set
or
Change between different data sets
End display of recordings
c) Clear manual recordings:

Press for 2 seconds
Select with or :
Clear nothing (cancel menu)
Clear all recordings
Clear the last recording

8.2 Automatic Recording with Selectable Cycle Time (“Func CYCL“)
the device will automatically record
measuring values at intervals of the set cycle time.
).
Storable data sets: 8000
A data set contains at slow measuring rate (rAtE SLo):
- Sensor 1: measuring value at the time of the recording
- Sensor 1: min-peak, max-peak since the last recording
- Sensor 2: measuring value at the time of the recording
- Sensor 2: min-peak, max-peak since the last recording
- Difference sensor 1-sensor 2: measuring value at the time of the recording
- CALC off *) Difference sensor 1-sensor 2: min-peak, max-peak since the last recording
CALC active *) Calculated value at the time of the recording and dummy value
At data set contains at fast measuring rate (rAtE FASt and rAtE P.dEt):
- Sensor 1: arithmetic average since the last recording
- Sensor 1: min-peak, max-peak since the last recording
- Sensor 2: arithmetic average since the last recording
- Sensor 2: min-peak, max-peak since the last recording
- Difference sensor 1-sensor 2: arithmetic average since the last recording
- CALC off *) Difference sensor 1-sensor 2: min-peak, max-peak since the last recording
CALC active *) Calculated value at the time of the recording and dummy value
*) CALC-Function: see chapter 7.6, 7.7