Instructions

H82.0.03.6C-08 Operating Manual GMH 5155 page 10 of 22
_____________________________________________________ _____________________________________________________________________________
Menu
Parameter
Value
Description
or
Set Logger: Settings for logger function
*
Selection of logger function
*
CYCL
Cyclic: cyclic logger
Stor
Store: manual recording
oFF
No logger activated
0:01 .. 60:00
Cycle time in [minutes: seconds] (for cyclic logger)
*
on/oFF
Low-power-logger with low power consumption
(only for cyclic logger and slow measuring rate)
*
Set Clock: Settings for real time clock
HH:MM
Clock: set time hours: minutes
YYYY
Year: set year
TT.MM
Date: set date day. month
rEAd Logg: Read calibration data:
p.r.t. chapter 8.1 Manual Recording (“Func-Stor”)
(*) If logger memory contains data sets parameters marked with (*) cannot be called. You
have to clear memory to change these parameters! (button 6, p.r.t. chapter 8)
(**) Parameter can only be called if corresponding sensor is plugged to connection 1. The
settings are applied to the sensor at connection 2 (if any connected).
7 Remarks to Special Features
7.1 Different Pressure Units
Depending on the connected sensors different units can be selected via menu (Unit). The measuring range
of the sensors may restrict the choice!
User-Unit
For pressure units, which are not covered by the ones printed on the display, there can be done a manual
-defi unit.
Example:
To enable a GMSD 2 BR - K51 ( -1000 2000 mbar) to display the unit kg/cm
2
, following settings (bold)
have to be entered:
kg/cm
2
Torr
atm
at
bASE user:
bar
mbar
bar
bar
DP user:
.0000
.0000
.0000
.0000
FACt user:
1.0197
.7433
0.9869
1.0197