H82.0.03.6C-08 Operating Manual GMH 5155 page 9 of 22
_____________________________________________________ _____________________________________________________________________________
Menu
Parameter
Value
Description
or
Set Output: Universal output adjustments
oFF
Interface and analog output off
SEr:
Serial interface activated
dAC:
Analog output activated
01, 11..91
Base address for serial interface communication
(only at Out = SEr)
CH 1, CH 2
or
DIF
CH
Channel that should be used for analog output
(only at Out = dAC)
sensor dep.
e.g.-5.00..5.00 mbar
Zero-point setting for Out = dAC: Measuring value which should
correspond to output 0 V
sensor dep.
e.g.-5.00..5.00 mbar
Slope setting for Out = dAC: Measuring value which should
correspond to output 1 V
Set Corr: Input adjustment
*
Zero adjustment/offset of sensors 1
oFF
No zero adjustment of sensors 1
Sensor dep.
e.g.-5.00..5.00 mbar
The offset of sensor 1 will be displaced by this value to compensate
for deviations of the probe or of the measuring device.
Slope adjustment of sensors 1
oFF
No slope adjustment of sensor 1
-2.000 …
2.000
The slope of sensor 1 will be changed by this factor [%] to
compensate deviations of probe or measuring device
Zero adjustment/offset of sensors 2
oFF
No zero adjustment of sensors 1
Sensor dep.
e.g.-5.00..5.00 mbar
The offset of sensor 2 will be displaced by this value to compensate
for deviations of the probe or of the measuring device.
Slope adjustment of sensors 2
oFF
No slope adjustment of sensor 2
-2.000 …
2.000
The slope of sensor 2 will be changed by this factor [%] to
compensate deviations of probe or measuring device
Set Alarm: Settings for alarm function
On/No.So
Alarm Sensor 1 on, with/without buzzer
OFF
No alarm function for Sensor 1
Sensor1-Min
... AL.1-Hi
Min-alarm-limit sensor 1 (not if AL.1 = oFF)
Sensor1-Min is lower display range limit of sensor 1
AL.1-Lo ...
Sensor1-Max
Max-alarm-limit sensor 1 (not if AL.1 = oFF)
Sensor1-Max is upper display range limit of sensor 1
On/No.So
Alarm Sensor 2 on, with/without buzzer
OFF
No alarm function for Sensor 2
Sensor2-Min
... AL.2-Hi
Min-alarm-limit sensor 2 (not if AL.2 = oFF)
Sensor2-Min is lower display range limit of sensor 2
AL.2-Lo ...
Sensor2-Max
Max-alarm-limit sensor 2 (not if AL.2 = oFF)
Sensor2-Max is upper display range limit of sensor 2
On/No.So
Alarm sensor difference on with/without buzzer
OFF
No alarm function for sensor difference
-19999 …
AL.DIF-Hi
Min-alarm-limit difference (not if AL.DIF = oFF)
AL.DIF-Lo …
19999