H82.0.02.6C-05 Operating Manual GMH 5150 page 9 of 22
_____________________________________________________ _____________________________________________________________________________
Menu
Parameter
Value
Description
or
Set Output: Universal output adjustments
oFF
Interface and analog output off
SEr:
Serial interface activated
dAC:
Analog output activated
01, 11..91
Base address for serial interface communication
(only at Out = SEr)
sensor dep.
e.g.-5.00..5.00 mbar
Zero-point setting for Out = dAC: Measuring value which should
correspond to output 0 V
sensor dep.
e.g.-5.00..5.00 mbar
Slope setting for Out = dAC: Measuring value which should
correspond to output 1 V
Set Corr: Input adjustment
*
Zero adjustment/offset of sensors
*
oFF
No zero adjustment of sensors
Sensor dep.
e.g.-5.00..5.00 mbar
The offset of sensor will be displaced by this value to compensate
for deviations of the probe or of the measuring device.
Slope adjustment of sensors
*
oFF
No slope adjustment of sensor
-2.000 …
2.000
The slope of sensor will be changed by this factor [%] to
compensate deviations of probe or measuring device
Set Alarm: Settings for alarm function
On/No.So
Alarm on, with/without buzzer
OFF
No alarm function
Sensor-Min
... AL.Hi
Min-alarm-limit sensor (not if AL. = oFF)
Sensor-Min is lower display range limit of sensor
AL.Lo ...
Sensor-Max
Max-alarm-limit sensor (not if AL. = oFF)
Sensor-Max is upper display range limit of sensor
Set Logger: Settings for logger function
*
Selection of logger function
*
CYCL
Cyclic: cyclic logger
Stor
Store: manual recording
oFF
No logger activated
0:01 .. 60:00
Cycle time in [minutes: seconds] (for cyclic logger)
*
on/oFF
Low-power-logger with low power consumption
(only for cyclic logger and slow measuring rate)
*
Set Clock: Settings for real time clock
HH:MM
Clock: set time hours: minutes
YYYY
Year: set year
TT.MM
Date: set date day. month
rEAd Logg: Read calibration data:
p.r.t. chapter 8.1 Manual Recording (“Func-Stor”)
(*) If logger memory contains data sets parameters marked with (*) cannot be called. You
have to clear memory to change these parameters! (button 6, p.r.t. chapter 8)
(**) Parameter can only be called if corresponding sensor is plugged.