E39.0.1X.6C-01 Data logger for as of version 1.0 Operating Instruction standard signals T-Logg 120 ... T-Logg 120 W - ... T-Logg 120 K - ... GREISINGER electronic GmbH D - 93128 Regenstauf, Hans-Sachs-Straße 26 ℡ +49 (0) 9402 / 9383-0 +49 (0) 9402 / 9383-33 | info@greisinger.
E39.0.1X.6C-01 1 Operating Instruction T-Logg 120 page 2 of 8 General The logger T-Logg 120 ... is designed as a cost-efficient solution for monitoring standard signals. It enables an individual programming of the recording time. The last 16,000 measuring values can be stored in the memory. In addition, the LCD-display indicates both the temperature measured at the moment and the operating status of the logger.
E39.0.1X.6C-01 4.2 Operating Instruction T-Logg 120 page 3 of 8 T-Logg 120 W - ... : Assignment of the angle-type plug Table of connections angle-type socket and jack Pin no. Inputs wire colour 0 ... X Volt 0 ... 20 mA 4 ... 20 mA 1 blue feeded through, signal + feeded through feeded through 2 red feeded through , signal - feeded through T-Logg 3 black feeded through T-Logg feeded through 4 yellow feeded through feeded through feeded through Input signal: 0 ...
E39.0.1X.6C-01 4.3 Operating Instruction T-Logg 120 page 4 of 8 T-Logg 120 K - ... : Assignment of the connection cable The connection cable has following assignment: 5 Wire colour Meaning red signal + white signal - Advice regarding state of logger upon delivery Upon its delivery the logger is in a kind of 'sleeping state': The display does not show anything, the power consumption is at its minimum. The T-Logg ‘wakes up’ as a communication link with a software has been established.
E39.0.1X.6C-01 Operating Instruction T-Logg 120 page 5 of 8 ALARM LOW: The measured value is below the min. alarm limit. ALARM HIGH: The measured value has exceeded the max. alarm limit. ERROR 1: The input signal has exceeded the measuring range of the logger. ERROR 2: The input signal has been fallen below the measuring range of the logger. ERROR 3: The max. display range (9999) of the logger is exceeded. ERROR 4: The min. display range (-1999) of the logger is under-run.
E39.0.1X.6C-01 9 Operating Instruction T-Logg 120 page 6 of 8 Battery replacement Notice: As soon as BAT appears at display, the battery needs to be replaced. If battery voltage power decreases continuously, data logging will be stopped (compare with operating status ‘HALT’). Available stored data remains in the memory and do not get lost. It’s possible to read the measuring data after replacement of the battery! Replacement: 1.
E39.0.1X.6C-01 Operating Instruction T-Logg 120 page 7 of 8 10 Specification Input signal: T-Logg 120 .. - 001: T-Logg 120 .. - 002: T-Logg 120 .. - 010: T-Logg 120 .. - 020: T-Logg 120 .. - 420: refer to device type, or in according to type plate 0 – 1 Volt input resistance = ~ 100 kOhm 0 – 2 Volt input resistance = ~ 200 kOhm 0 – 10 Volt input resistance = ~ 500 kOhm 0 – 20 mA shunt resistance = ~ 70 Ohm 4 – 20 mA shunt resistance = ~ 70 Ohm Display range: Scaling: Decimal point: -1999 ...
E39.0.1X.