User manual
E40.0.01.6C-02 Operating Manual EASY
LOG
 80 CL  Page 8 of 12 
9. Operating displays 
The EASY
LOG
 provides two LCD-displays:  7 mm   (main display) and 
  5.8 mm  (secondary display). 
9.1 Measured value display 
The LCD mainly displays the measured values:
a) 
°C
%
 AL   Logg 
b) 
hPa
°C
 SL Logg 
MEASURED VALUE DISPLAY 
Alternating display of the following measuring channels: 
a) main display:  humidity [ % ] 
  sec. display:  temperature [ °C ]  
b) main display:  selected calculated variable with the  
      corresponding unit arrow (here: Td) 
  sec. display:  air pressure [ hPa ] 
The particular measuring channels are displayed for about 4 sec. 
and then are automatically switched to the next ones.  
• The arrow “Logg“ signals that the logger is recording. 
• The arrow “AL“ signals an alarm state at minimum one channel.
• The arrow “SL“ signals an active sea-level correction . 
9.2  Logger status messages 
At stand-alone-operation (= no EASYBus connected) logger status messages are dis-
played. 
STOP: 
The EASY
LOG
 recording has been ›stopped‹. 
The logger memory is empty.
Note:  At this state the power consumption is at its 
minimum.  
HOLD: 
The EASY
LOG
 recording has been ›halted‹. No data is stored, 
but the logger memory contains data. 
The number of the saved recording sets is shown at the 
secondary display. 
START DELAY: 
The logger is activated, but no data is recorded yet. 
As soon as the start delay is run out the logger begins with the 
recording according to the programmed starting condition.  
START ALARM: 
The logger is activated, but no data is recorded yet. 
The logger begins with the recording as soon as the measured  
values are within the selected min- and max-alarm boundaries. 