User manual
E39.0.31.6C-02  Operating instruction T-Logg 160  page 6 of 8 
7.  Insert the new button cell (type CR 2032) with correct 
polarity in the socket (the positive pole is on the retaining 
bracket). 
8.  Induct the sensor connection into the hole of the probe 
tube and put the board into the housing.  
9.  Thread the foam rubber onto the connector again and 
move it back into the probe tube. 
10.  Plug the sensor element to the connector (mind right 
mounting position). 
 Note:  The connection pins can break if they are 
inappropriately handled! 
11.  Screw the protective head on the sensor tube. 
12.  Reassemble the cover to the housing. 
Finished 
Environmental Reference!  Empty and defective batteries must not be disposed in the 
  regular domestic waste. 
 Return the used batteries to an authorised battery 
 collecting point or send the batteries directly to us 
 (sufficiently stamped). 