User manual
G14.0.11.6C-04 Operating Manual MINISOFT page 5
GREISINGER electronic GmbH Hans-Sachs-Straße 26 Fon: 0049 9402 / 9388-0
D-93128 Regenstauf Fax.: 0049 9402 / 9383-33
4.4 Settings
Display of all the editable properties and settings of the logger.
Please note: If the logger contains recorded data the settings can only be changed if the data are deleted!!
The Window Title can be edited here
Identification (up to 16 digits, older logger types do not support an identification, ‚-
--‘ will be displayed then)
The identification is displayed at the top of the logger window and will help you to
identify the logger. We, therefore, recommend to choose an identification with
regard to the logger position, e.g. ‘cooling area’ or ‘climate room 1’.
Alarm- Min, -Max, -Delay:
input of alarm delay as a multiple of a minute, older logger types do not support an
alarm delay, the display then is always 0 min.
Legend: description regarding settings (only available if menue ‘Logger – Show
Legends’ has been activated).
4.5 About
Display of all the remaining properties and settings of the logger.
Address: cannot be changed via MINISOFT or is static at a T-Logg
(To set the address of a MINILOG please use our EBxKonfig software)
Display data
•  Range
•  Type of measurement: older logger types do not support a type of
measurement display, ‚---‘ will be displayed then.
Correction: if supported by logger
The Version number of the logger
5  Start of the Recordings And Starting Conditions
When starting a new logger recording you have a choice between various starting conditions:
Please note: The logger time will be set to computer time after every new start.
Application Examples:
Start a recording at a certain point of time:
Set logger, start new logger recording via the option ‘immediately‘ and the point of time desired.
The logger display alternates between the measuring value and ‚St.dE‘ (= start with delay). Disconnect logger and take to place of application.
Recording starts as soon as the point of time has been reached, ‚St.dE‘ disappears.
Place of logger application not in the vicinity of a computer, recording to start at place of application:
Set logger at computer, start new logger recording series via the option ‘when externally started’ The logger display alternates between the
measuring value and ‚St.Et‘ (= wait for ‘external start’). Disconnect logger and start at the place of application by pressing the button of the starting
key ‚EBSK-1‘: Logger recording starts, ‚St.Et‘ disappears.
Application is not in the vicinity of a computer but in a cooling area with ~ –10°C. Recording to start after logger has been placed there.
Set max. alarm limit of logger to 0°C; set min. alarm limit to –20°C. Start new logger recording series via the option ‘when alarm is cleared’. At room
temperature (>0°C !) the logger will be at max. alarm (logger window: register ‘Device’ – ‘Message’); the display alternates between the measuring
value and ‚St.Al‘ (= wait for ‚start alarm‘). Disconnect logger and place it in cooling area.
As soon as the logger temperature in the cooling area falls below 0°C the alarm will be cleared and the logger starts recording, ‚St.Al‘ disappears
(no computer required at the place of application).
Start at a certain point of time or
immediately:
Logger recording starts as soon as the
time set has been reached. If the time is
not changed the loggers selected will start
immediately. (this selection is sufficient for
most applications)
Start when alarm is cleared: the logger
recording will only start if all alarms have
been cleared even if the time set has
already been reached.
Alarm is given eg. if the meas. value is
smaller than the min. alarm limit set
(
lo
gg
er window re
g
ister 
‚
Extras‘
)
.
Start when externally started: the logger recording will only
be started externally, even if the time set has been reached.
An external start can be carried out on site with the help of
the ‚EBSK-1‘ starting key. 
(not available for T-Logg...)