User manual
G14.0.11.6C-04 Operating Manual MINISOFT page 3
GREISINGER electronic GmbH Hans-Sachs-Straße 26 Fon: 0049 9402 / 9388-0
D-93128 Regenstauf Fax.: 0049 9402 / 9383-33
Step 3: Start New Logger Recording
To start a new logger recording select ‘Start New Logger Recording’ from the ‘Logger’ menue. (Unless you have
already done so: link to connected loggers, see above).
Note: Data that have already been recorded will be lost after a new logger recording has been started!
Read the logger data and save them into a file, if necessary! (p.r.t. step 6)
Alarm limits and identification will have to be entered prior to the start (p.r.t. step 2). Upon logger
start the logger time will automatically be updated to the current time.
  Further description: chapter 5
Step 4: Read Logger Data
Link to connected loggers unless you have already done so (see above).
To read out data of a logger recording select ‘Read Logger Data’ from ‘Logger’ menue; the relevant window will be
opened. If more loggers are connected that may contain data, you can either read ‘all data’ of the loggers selected or
select a common period of time.
The data that have been read out will be displayed in the register ‘Read Data’ in the table.
Step 5: Stop Logger Recording
If the logger is not used for a longer period of time the battery service life can be extrended by stopping the logger.
Once stopped the logger is only measuring if connected to a level converter (logger supply via level converter). To stop
logger select ‘Stop Logger Recording’ from the ‘Logger’ menue (unless you have already done so: link logger, see
above).
Please note: Older MINILOG versions will loose their recorded data as soon as logger is stopped !
Step 6: File Operation: Saving And Loading
It is possible to attach a comment (logger window register ‘Comment’) to the data read in via the logger so that they
can be stored in data files: menue ‘File – Save’ or ‘File – Save As’.
Data that have been saved can be displayed again by means of the ‘File Open’ command.
Attention: Do not mix up with ‘File – Save View’ or ‘Open View’ – p.r.t. chapter 9.
Hint:  The files saved at last will be displayed in the menue ‘File’ and can be opened by means of a
mouseclick.
Select logger to be started
Enter time as of which logger is to be started
(p.r.t. chapter 5)
Selection of starting conditions
(p.r.t. chapter 5)
Type of memory
(‚endless: ..‘: permanent recording, if memory is
full, the data at the beginn-ing of the memory will
be overwritten)
Interval: 2s...5h(depending on logger).
ex..: ‚15 min‘ -> logger recording measuring value
at 15 min intervals