3.1 „Func-Stor“: Storing Single Measurements
Each time when „store“ (key 6) is pressed a measurement and its time stamp will be recorded.
The recorded data can be viewed either in the display (when calling the configuration an additional menu
„REAd LoGG“ is displayed, see below) or by means of the interface and a PC with GSOFT3050-software.
Please note: For the Func-Stor-logging function it is not allowed to change the pressure sensor after values
are stored, otherwise invalid data could be read out.
For the read out of the logger the sensor has to be kept connected!
Max. number of measurings: 99
A measuring contains:
- sensor 1, current measuring value at the time of recording
- sensor 1, min peak since the last recording
- sensor 1, max peak since the last recording
- sensor 2, current measuring value at the time of recording
- sensor 2, min peak since the last recording
- sensor 2, max peak since the last recording
- difference sensor 1 - sensor 2, current measuring value at time of recording
- difference sensor 1 - sensor 2, min peak since the last recording
- difference sensor 1 - sensor 2, max peak since the last recording
- time and date of the recording
After each recording „St. XX“ will be displayed for a short time. XX represents the number of the recording.
If logger memory contains recordings already:
When „Store“ is pressed for 2 seconds, the choice for clearing the logger memory will be displayed:
The selection can be made by 5 (key 2) and 6 (key 5). "Quit" (key 6) enters the choice.
If the logger memory is full, the display will show:
Viewing Recorded Measurings
Within the „LoGG Stor“ function the measurings can be viewed directly in the display not only by means of a
computer (like at „Func CYCL“): press 2 seconds „Set“ (key 4): The first menu displayed now is „rEAd LoGG
(read logger data). After pressing 4 (key 3) the measurement recorded last will be displayed, changing
between the different values referring to the measurement also is done by pressing 4.
Changing the measurement is done by pressing the keys 5 or 6.
3.2 „Func-CYCL“: Automatic Recording With Selectable Logger-Cycle-Time
The Logger-Cycle-Time is settable (p.r.t. Configuration). For example „CYCL“ = 60: A measuring is recorded
after each 60 seconds.
When the slow measurement "rAtE-Slo" is chosen, additionally a low power function is available: „Lo.Po“.
If „Lo.Po“ is on, the device only will take a measurement at the point of time of the recording. In between the
recordings the measuring shuts down. This decreases the power consumption enormously and therefore is
recommended e.g. for long time recordings where no mains adapter is available.
Max. number of measurings: 4000 (at max. 64 recording sequences)
Cycle time: 1...3600 seconds (=1h), selectable in the configuration
A measuring contains:
! slow measuring rate (rAtE Slo):
- sensor 1, current measuring value at the time of recording
- sensor 1, min peak, max peak since the last recording
- sensor 2, current measuring value at the time of recording
- sensor 2, min peak, max peak since the last recording
- difference sensor 1 – sensor 2, current measuring value at time of recording
- difference sensor 1 – sensor 2, min peak, max peak since the last recording
Clear all
recordings
Clear the last
recording
Clear nothing
(cancel menu)