User's Guide

83
LDX 100
User’s Guide
1 On the CGA, go to the Color menu > Color Temperature and select one of the
memory positions (FL50, FL60, AW1 or AW2) in which to store the measured color
temperature value.
On the camera use the color temperature switch to select one of the memory positions
(FL50, FL60, AW1 or AW2) in which to store the measured color temperature value.
2 In the Color Temperature menu on the CGA, tap the AUTO WHITE button to start the
Auto White balance procedure.
On the camera, press and hold (2 sec) the WHITE BALANCE button on the front panel to
start the Auto White Balance procedure.
3 The following appears in the viewfinder:
Note
Note
If you did not select one of the memory positions FL50, FL60, AW1 or AW2, a message (“Set
Color Temp”) appears in the viewfinder.
4 Point the camera so that the reference white surface is between the two small black
boxes.
5 On the CGA, tap the WHITE BALANCE button again.
On the camera, press the WHITE BALANCE button on the front panel again to start the
measurement procedure.
6 A message indicating that the process is running appears:
Note
Note
If there is insufficient light, the “light level too low” message appears in the viewfinder.
AWHITE: window
AWHITE: running