6.0 Technical specifications
6.1 Operating conditions
Ambient temperature range
5 to 40°C
Altitude above sea level
Up to 2,000m (6,500ft)
Operating environment
Indoor use only
Maximum relative humidity
80% RH up to 31°C decreasing to 50% RH at 40°C
6.2 Electrical details
Mains supply: 220-240V @ 50/60Hz or 110-120V @ 50/60Hz
Pollution degree: 2
Installation category: II
Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
6.3 Product performance
TX150
TX150L
TXF200
TXF200L
Settable temperature range
-50°C to 150°C
-50°C to 200°C
Stability (DIN 12876)
±0.01°C
Uniformity (DIN 12876)
±0.05°C
Max pump head pressure
310mBar
530mBar
Max pump flow rate
18l/min
22l/min (adjustable)
Max current consumption
9A
12.5A
9A
12.5A
Heater power
1.8kW
1.4kW
1.8kW
1.4kW
6.4 Bath accessories information
ST5
ST12
ST18
ST26
ST38
Tank capacity (litres)
5
12
18
26
38
Liquid depth min/max (mm)
85/140
85/140
75/130
125/180
125/180
P5
P12
P18
Tank capacity (litres)
5
12
18
Liquid depth min/max (mm)*
85/140
85/140
85/140