Operating Manual Manual

30424 V6 TX150 & TXF200
Page 11 Operating Manual
www.grantinstruments.com
Pumping heat transfer liquid around an external system can lead to hazards that are
outside the control of Grant Instruments. It is essential that the user conducts a risk
assessment of the entire equipment installation to ensure that correctly rated materials
have been used throughout and that the system can be used safely.
5.1.5 Using the switch over relay output
An internal relay provides switch over contacts that can be used to control external
equipment. The pin connections on the rear panel 3 pin circular connector are:
For the connecting cable use a mating XLR style connector such as the NC3FXX
manufactured by Neutrik AG
The switch over contacts are rated at 24V AC or DC 2A maximum.
3
2
1
Normally Closed
Normally Open
Common
Rear
Panel
connector
Relay
B
A