Operating Manual Manual

30424 V6 TX150 & TXF200
Page 9 Operating Manual
www.grantinstruments.com
Temp
range
Recommended
liquid
Cut-out
setting
Comments
-50°C to
150°C or
200°C
None
As
required
for
safety
Override range - The user must select a
suitable safe liquid and carry out their own
risk assessment before use. Note that the
control when using this setting may be
affected as the liquid characteristics are
unknown.
4.7 Installation
4.8 Electrical supply
4.9 Using accessory cooling (C1G, C2G, CW5)
Accessory cooling is required for operation at temperatures below ambient. Refrigerated
dip coolers (C1G and C2G) can be used for operation down to -15°C. A water heat
exchanger coil (CW5) can be used for operation at or around ambient. The coils can be
fitted:
1. Attach u-shaped coil locating rod to cover plate using two fixings supplied (1).
2. Fit the cooling coil through the hole in the bridge plate and align outlet pipes with cut-
out notch (2).
3. Fit cover plate (3) onto bridge plate and attach with fixings supplied. The coil locating
rod should press fit against the inside of the coil and hold it rigidly. Ensure the coil is
held safely and securely before operation.
2
1
3
CW5
C1G & C2G
2
1
3