The Sphygmomanometer (MD6000) is intended for use in the electromagnetic environment specified below. The customer
or the user of the Sphygmomanometer (MD6000) should assure that it is used in such an environment.
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst
IEC 61000-4-4
Power supply lines: ±2 kV Power supply lines: ±2 kV
Surge
IEC 61000-4-5
line(s) to line(s): ±1 kV
100 kHz repetition frequency
line(s) to line(s): ±1 kV
100 kHz repetition frequency
Power frequency magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mians voltage prior to application of the test level.
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz - 2,7 GHz
80% AM at 1 kHz
10 V/m
80 MHz - 2,7 GHz
80% AM at 1 kHz
Conduced RF
IEC 61000-4-6
150KHz to 80MHz:
3Vrms
6Vrms (in ISM and amateur radio
bands) 80% Am at 1kHz
150KHz to 80MHz:
3Vrms
6Vrms (in ISM and amateur radio
bands) 80% Am at 1kHz
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
0% 0.5 cycle
At 0˚, 45˚, 90˚, 135˚, 180˚, 225˚, 270˚ and
315˚ 0% 1 cycle And 70% 25/30 cycles
Single phase: at 0
0% 300 cycle
0% 0.5 cycle
At 0˚, 45˚, 90˚, 135˚, 180˚, 225˚, 270˚ and
315˚
0% 1 cycle And 70% 25/30 cycles
Single phase: at 0
0% 300 cycle
Immunity test IEC 60601-1-2 Test level Compliance level
Guidance and manufacturer’s declaration - electromagnetic Immunity