User's Manual
Table Of Contents
Appendix III
Test Frequency
(MHz)
385 380-390 TETRA 400 1,8 0.3 27Pulse modulation
18Hz
450 430-470 GMRS 460,
FRS 460
2 0.3 28
2 0.3 28
FM ± 5 kHz
deviation 1 kHz sine
2450 2400-2570 Bluetooth, WLAN,
802.11 b/g/n, RFID
2450, LTE Band 7
2 0.3 28Pulse modulation
217 Hz
710
745
780
810
870
930
704-787 LTE Band 13, 17
800-960 GSM 800/900, TETRA
800, iDEN 820, CDMA
850, LTE Band 5
0,2 0.3 9Pulse modulation
217 Hz
Pulse modulation
18 Hz
0,2 0.3 9
5240
5500
5785
5100-5800 WLAN 802.11 a/n
Pulse modulation
217 Hz
2 0.3 28
1720
1845
1970
1700-1990 GSM 1800; CDMA
1900; GSM 1900;
DECT; LTE Band 1,
3, 4, 25; UMTS
Pulse modulation
217 Hz
Radiated RF
IEC61000-4-3
(Test
specifications
for ENCLOSURE
PORT
IMMUNITY to
RF wireless
communica-
tions
equipment)
Band
(MHz)
Service Modulation Modulation
(W)
Distance
(m)
IMMUNITY
TEST LEVEL(V/m)
Guidance and manufacturer’s declaration - electromagnetic Immunity
The Sphygmomanometer (MD5900/MD5910/MD5920/MD5930/MD5940/MD5950/MD5960/MD5970/MD5980/MD5990) is intended
for use in the electromagnetic environment specified below. The customer or the user of the
Sphygmomanometer (MD5900/
MD5910/MD5920/MD5930/MD5940/MD5950/MD5960/MD5970/MD5980/MD5990) should assure that it is used in such an
environment.