User's Manual
Table Of Contents
Appendix II
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst
IEC 61000-4-4
Power supply lines: ±2 kV Power supply lines: ±2 kV
Surge
IEC 61000-4-5
line(s) to line(s): ±1 kV 100 kHz
repetition frequency
line(s) to line(s): ±1 kV 100 kHz
repetition frequency
Power frequency magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mians voltage prior to application of the test level.
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz - 2,7 GHz
80% AM at 1 kHz
10 V/m
80 MHz - 2,7 GHz
80% AM at 1 kHz
Conduced RF
IEC 61000-4-6
150KHz to 80MHz:
3Vrms
6Vrms (in ISM and amateur radio bands)
80% Am at 1kHz
150KHz to 80MHz:
3Vrms
6Vrms (in ISM and amateur radio bands)
80% Am at 1kHz
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
0% 0.5 cycle
At 0˚, 45˚, 90˚, 135˚, 180˚, 225˚, 270˚ and 315˚ 0%
1 cycle And 70% 25/30 cycles
Single phase: at 0
0% 300 cycle
0% 0.5 cycle
At 0˚, 45˚, 90˚, 135˚, 180˚, 225˚, 270˚ and 315˚
0% 1 cycle And 70% 25/30 cycles
Single phase: at 0
0% 300 cycle
Immunity test IEC 60601-1-2 Test level
Compliance level
Guidance and manufacturer’s declaration - electromagnetic Immunity
The Sphygmomanometer (MD5900/MD5910/MD5920/MD5930/MD5940/MD5950/MD5960/MD5970/MD5980/MD5990) is intended
for use in the electromagnetic environment specified below. The customer or the user of the Sphygmomanometer (MD5900/
MD5910/MD5920/MD5930/MD5940/MD5950/MD5960/MD5970/MD5980/MD5990) should assure that it is used in such an
environment.