6  OPERATING INSTRUCTIONS
LF1041 ONLY: EMERGENCY PROCEDURES
Emergency stop: The control box is equipped with a red 
emergency stop button. This button, also known as the 
control box power switch, is to be used in the event the hand 
control pendant ceases to function correctly. To engage, hit 
the button rmly with the palm of your hand. The button will 
lock in the depressed position, stopping all motor function. 
To disengage, turn the button clockwise; it will pop out and 
return to the ON position.
Emergency lowering: The motor is equipped with an 
emergency release, which is located on top of the motor 
as shown in the picture at right. To engage, pull up on the 
release, which will disengage the gearing in the cylinder, and 
lower the patient. The emergency release is adjustable by 
turning the screw adjacent to the pull stem. Turn the screw 
clockwise to increase and counter-clockwise to decrease the 
sensitivity of the release mechanism.
  WARNING: Use caution when using the emergency 
release. Pull back slowly on the release, which will 
gradually lower the patient. Pulling back too quickly 
could cause the patient to be lowered too rapidly, 
which could result in serious personal injury. Do 
not, under any circumstances, use the emergency 
release for lowering the patient when the hand control 
pendant and battery are operating correctly. The 
emergency release should not be used in lieu of the 
hand control pendant and battery.
motor
adjustment
 screw
emergency
release
emergency
stop button
emergency stop button
emergency release