User's Manual
16  John Bunn O2 Premier Oxygen Concentrator
6  TROUBLESHOOTING GUIDE
 SYMPTOM  CHECK / ACTION
Concentrator power is  Check if concentrator is unplugged; plug in. 
on and alarm is on, but 
concentrator doesn’t  Check if circuit breaker is tripped; x circuit 
work  breaker. 
  If concentrator still doesn’t work, contact 
  Graham-Field equipment provider.
 SYMPTOM  CHECK / ACTION
Concentrator works,  Inspect tubes and cannula to ensure that they are 
but little or no oxygen  not kinked or blocked. Smooth tubes to unblock.
from outlet 
  Check humidier bottle, if so equipped, for leaks 
  or jamming. Give humidier bottle a slight shake 
  to open safety valve. Clean humidier bottle, 
  reinstall and tighten.
  If condition persists, contact Graham-Field 
  equipment provider.
 SYMPTOM  CHECK / ACTION
Flowmeter ball located  Adjust owmeter knob. If condition persists, 
on “0” position or oat  contact Graham-Field equipment provider. 
is unstable
 SYMPTOM  CHECK / ACTION
Concentrator works,  If yellow light blinks, and alarm sounds at low 
but makes abnormal  intervals, or yellow light illuminates continuously 
sound  and alarm sounds, contact Graham-Field 
  equipment provider.
 SYMPTOM  CHECK / ACTION
Concentrator works,  OK to continue use, but contact Graham-Field 
but yellow fault light  equipment provider immediately. Concentrator’s 
lluminates  oxygen concentration is safe, but decreasing. 
(JB0160-015B,  When oxygen concentration is too low for safe 
 JB0160-015,  use, concentrator will shut off and optional O
2
 JB0160-015B-220,  sensor red warning light will illuminate. 
 JB0160-015-220 only)