User Guide
760000-INS-LAB-RevA08  11
Note: The mattress system will automatically enter Autofirm 
mode for the quickest inflation at set-up; initial inflation takes 
approximately twenty minutes.
3.  The pump will begin delivering air into the mattress.
4.  When the selected pressure level is reached, the pump 
will enter alternating mode. The mattress can then be 
adjusted to the desired pressure.
5.  Replace all sheets, blankets, and pillows on the bed.
pressure adjustment
Generally, a lighter patient will need a lower (softer) set-
ting while a heavier patient will need a higher (rmer) 
setting, but pressure adjustment must ultimately be based 
on the patient's weight distribution. A short, stocky person 
may be the same weight as a tall, thin person, but their 
settings will not be the same. The mattress should be rm 
enough to support the patient, yet soft enough so the 
patient will sink down to eliminate pressure points. Initially 
inate the mattress to maximum pressure. Then, when 
patient lies on the mattress, test the pressure by placing 
four ngers below the mattress under the patient's but-
tock area; your ngers should just t. If the t is too loose, 
decrease the setting. If your ngers do not t well, increase 
the setting two notches rmer. Wait ten minutes and then 
check again. You should only have to adjust this setting 
one time per patient. Please consult your physician for an 
appropriate setting.
quICk dIsConneCt
Pull the power plug from the wall connector to disconnect 
the device quickly.