Product Manual
4077-INS-LAB-RevB09 6
WARNING: The alarm must be congured to different situations of individual pa-
tients. Ensure that audio sounds can be activated when alarm occurs.
WARNING: DO NOT switch on device power until all cables have been properly
connected and veried.
WARNING: DO NOT touch the signal input or output connector and the patient si-
multaneously.
WARNING: Accessory equipment connected to the analog and digital interfaces
must be certied according to the respective IEC standards (e.g. IEC 950 for data
processing equipment and IEC 60601-1 for medical equipment). Furthermore all
congurations shall comply with the valid version of the system standard IEC
60601-1-1. Anyone who connects additional equipment to the signal input con-
nector or signal output connector congures a medical system, and is therefore
responsible that the system complies with the requirements of the valid version of
the system standard IEC 60601-1-1. If in doubt, consult your Graham-Field equip-
ment provider.
WARNING: Electromagnetic Interference—Ensure that the environment in which
the Fetal Monitor is installed is not subject to any source of strong electromagnet-
ic interference, such as radio transmitters, mobile telephones, etc. The device will
show a xed gure at the LCD under the environment of strong electromagnetic
interference, and CANNOT be used as a diagnostic criterion.
CAUTION:
To reduce the risk of minor personal injury and/or product/property damage
Caution: The device is designed for continuous use and is “ordinary” (i.e. not drip
or splash-proof).
Caution: Keep the Fetal Monitor's environment clean and dust-free. Protect the
device from vibration, corrosive medicine, high temperatures and humidity.
Caution: When installing the Fetal Monitor in a cabinet, allow for adequate ventila-
tion, accessibility for servicing, and room for adequate visualization and operation.
Caution: Do not operate the Fetal Monitor if it is damp or wet because of condensa-
tion or spills. When moving the Fetal Monitor from a cold environment to a warm,
humid location, do not use the device until it has reached room temperature.
Caution: Do not immerse transducers in liquid. When using solutions, use sterile
wipes to avoid pouring fluids directly on the transducers.
Caution: Do not autoclave or gas sterilize the monitor or any accessory.
Caution: Do not dispose of this device with household waste. Dispose of this device
in accordance with your local laws and regulations.