Product Manual

4077-INS-LAB-RevB09 37
MONITOR continued
Ultrasound
Technique: Peak-peak detection technique and autocorrelation
Pulse Repetition Rate: 3.2KHz
Pulse Duration: 114µs
Nominal Frequency: 2.0MHz
Ultrasound Frequency: 2.0MHz±10%
P- <1 MPa
Iob<10 mW/cm2
Ispta<100 mW/cm2
FHR Range: 50bpm-210bpm
Resolution: 1bpm
Accuracy: ±2bpm
Earth Leakage Current: <10 uA @ 264 VAC applied to transducer
Dielectric Strength: >4000Vrms
Contraction
(External
TOCO)
TOCO Range: 0% - 100%, 135g strength corresponding to 100%
Sensitivity: 3.7uV/V/mmHg
Non-linear Error: ≤ ±10%
Resolution: 1%
Zero Mode: Automatic / Manual
Earth Leakage Current: <10 uA @ 264 VAC applied to transducer
Dielectric Strength: >4000 Vrms
AFM Technique: Pulsed Doppler Ultrasound
Range: 0-100 (%)
Resolution: 1%
Gain Control: Manual / Automatic
Marking Manual fetal movement mark