8 MAINTENANCE, CARE AND CLEANING
WARNING: Failure on the part of the responsible facility employing the use of this
equipment to implement a satisfactory maintenance schedule may cause undue
equipment failure and possible health hazard.
PREVENTIVE MAINTENANCE
Visual Inspection
Perform a visual inspection to ensure that the equipment, cables and transducers have
no visible evidence of damage before use. Thereafter, perform such visual inspection at
least once weekly. If damage is evident, replace the damaged component before use.
Routine Inspection
Perform periodic safety testing to ensure proper patient isolation from leakage currents,
including leakage current measurement and insulation testing, annually or as specied in
your facility’s test and inspection protocol.
Mechanical Inspection
1. Ensure that all exposed screws are securely fastened.
2. Check external cables for splits, cracks, or signs of twisting.
3. Replace any cable that shows serious damage.
4. Pay particular attention to the supply socket.
CARE AND CLEANING OF MONITOR
WARNING: Unplug the monitor from the AC power source and detach all accesso-
ries before cleaning. Do not immerse the unit in water or allow liquids to enter the
casing.
Caution: Keep the exterior surface of the monitor clean and free of dust and dirt.
Caution: Clean the monitor casing and display screen regularly. Use only non-caus-
tic detergents such as soap and water to clean the monitor casing.
Caution: Take extra care when cleaning the display surface, which is more sensi-
tive to rough handling, scratches and breakage than the other external surfaces of
the monitor. Use a clean, dry, soft cloth to wipe it gently.
Caution: Although the monitor is chemically resistant to most common hospital
cleansers and non-caustic detergents, different cleansers than those specified here
are not recommended and may stain the monitor.
Caution: Many cleansers must be diluted before use. Follow the manufacturer’s
directions carefully to avoid damaging the monitor.
Caution: Do not use strong solvents, such as acetone.
Caution: Never use an abrasive such as steel wool or metal polish.
Caution: Do not allow any liquid to enter the product, and do not immerse any
part of the monitor into any liquid.