6 SETUP
FETAL MONITOR CONNECTIONS
Connecting the Power Cable
1. Ensure the AC power supply of the you will be plugging the monitor into complies
with the following specication: AC 100V-240V, 50/60 Hz.
2. Plug one end of the power cable provided with the Fetal Monitor to the monitor
power socket (on monitor's right side).
3. Connect the other end of the power cable to a grounded 3-phase hospital-grade
power output.
FETAL MONITOR PRINTER PAPER
Feeding the Record Paper into the Fetal Monitor Recorder
If the paper runs out, feed paper into the recorder as follows:
1. Push the position simultaneously on both
sides of the recorder cover to open it.
2. Remove the “Z” type thermosensitive record
paper's wrapper. Position the green safety
band to the left and the face of the paper
downward so that the blank side—that with-
out print scale—faces the user.
3. Feed the record paper into the slot of the recorder (see picture at right). The paper
will exit from the notch automatically. If required, adjust the paper length with the
gear beside the handle.
4. If the paper is crooked, pull the handle up and push the gear to force the paper out.
Push the handle down and feed the paper again.
5. After closing the cover, make sure the paper can easily exit from the paper notch.
6. Guide the paper into the print head so that the paper aligns with the guide notch on
both sides.
Caution: Use care when inserting paper. Avoid damaging the thermosensitive print
head. Unless installing paper or fixing paper jam, do not leave the recorder door open.
Caution: Only use Graham-Field-approved paper to avoid poor printing quality,
deflection, or paper jam.
Note: When feeding paper, the black handle must be down. If paper jams, pull up the
handle first, push the gear to force the paper out, then feed the paper again.
Note: The paper exiting the notch must be aligned; otherwise, the data will be
inaccurate or paper may jam. There is a pull-paper bar within the recorder which can
be used to pull out jammed paper easily.
Note: The printing function cannot be executed when the ultrasound transducer
disconnects from the Fetal Monitor.
feeding paper
paper advance
slot
the blank side
(without print
scale) faces
user