FHR Alert
• Enabled:Enable/DisableaudioalarmwhenFHRexceedsthealarmlimit;ONmeans
enabled, OFF means disabled, and the default is ON.
• Lower:ThelowerlimitofacceptableFHRrange;Theadjustablerangeis50bpm-
130bpm, the default is 120bpm, and the increment is 5bpm.
• Upper:TheupperlimitofacceptableFHRrange;Theadjustablerangeis150bpm-
210bpm, the default is 160bpm, and the increment is 5bpm.
• Keep:Durationoftimeuntilalarmoccursifultrasoundsignalqualityisunacceptable;
The adjustable range is 0-30 seconds, the default is 30 seconds, and the increment is
5 seconds.
• Probe:Enable/Disableaudioalarmwhenultrasoundtransducer(US1)fallsoff;ON
means enabled, OFF means disabled, and the default is OFF.
• Quality:Enable/DisableaudioalarmwhenFHRsignalquality isunacceptable;ON
means enabled, OFF means disabled, and the default is OFF.
UA alert menu
• UAProbe:Enable/DisableaudioalarmwhenTOCOtransducerfallsoff;ONmeans
enabled, OFF means disabled, and the default is OFF.