Product Manual
4077-INS-LAB-RevB09 13
Trend/Menu Frame
Example of trend (left picture below) and menu (right picture below).
trendexample
FHRtrend
AFMtrend
TOCOtrend
menu example
Under auto monitoring and load status, the trend frame may display three trends at most
(see above left picture): FHR (US1) trend, AFM trend, and TOCO trend.
Two horizontal graticules are included at the FHR trend frame to make it easier for the
caregiver to observe heart rate trend or heart rates that exceed limits. The range is
120bpm -160bpm.
The y-axis of FHR trend indicates FHR values, and the range is 30bpm - 240bpm.
The y-axis of TOCO trend indicates TOCO values, and the range is 0% - 100%. There is a
horizontal baseline at the TOCO trend.
The AFM trend is displayed within the range of 30 - 60 units of FHR trend, and its y-axis
indicates AFM values.
The menu size on the screen is xed, and always occupies the trend frame, which makes
the trends temporarily invisible. See the MENU Section, for a detailed introduction to the
menu.
Note: The AFM trend is only for reference; please refer to remote mark.
Parameter Frame
The parameter frame (see picture below) is at the right of the trend frame. The param-
eters in the parameter frame may include: FHR (US1), TOCO, and FM count.
parameter
frame
147:FHRvalueofUS1
FHRrefreshmentrateofUS1
28:CurrentUAvalue
0:Manual/AutomaticFMcount