5. Dispense about 300–500 cc of material into a
graduated cylinder. The amount the system
measures will display in the measured volume
field
.
6. Press
to end th
e calibration.
7. Press
to get to the dispensed volume field
, then press to enter the field. Enter
the amount of material in the cylinder.
8. After the volume is entered, the system calculates
the new k-factor
and shows it on Setup
Screen 4.
NOTE: To clear the counter and begin the
calibration again, press
, move briefly to
another screen, then return to Setup Screen 4
and start over. If you press
without leaving
the screen, the counter will continue from where
it is, without clearing.
9. Press
to exit the screen.
10. Press
to exit Setup Mode.
Setting Modbus Address
See S
etup Screen 5. By default, the Modbus is
set t
oOff
. If you need the Modbus, set the
Modb
us mode to SLAVE
. The address
val
ue is between 1 and 247. The modbus address
cor
responds to the address of the Informer. See
App
endix A for further information.
Update Softwa
re
Software upda
tes are installed using a software
token (PN 16P
468), which is sent automatically when
a new version
of the software is released. Manual
3A1244 will a
ccompany any necessary software
updates. Fo
llow all instructions and warnings in
Manual 3A12
44 to update your Informer software.