User's Manual

Operation
Meter Calibra
tion
NOTE: See SetupScreen4for further screen
information, if needed.
When to Calib
rate
The first time the system is operated.
Whenever new materials are used in the system,
especially if the materials have viscosities that
differ significantly.
As part of regular maintenance to retain meter
accuracy.
Whenever a flow meter is serviced or replaced.
Read Before Calibration
Meter k-fa
ctor on Setup Screen 4 is updated
automati
cally after the calibration procedure is
completed. You also may manually edit the
k-factor if desired.
All values on this screen are in cc or cc/pulse,
regardless of the units set in the other Setup
screens.
Before calibrating the meter, be sure the system
is primed with material.
Disable alarms before calibration.
Calibration Steps
1. Press
to enter Setup Mode.
2. Press
to move to Setup Screen 4.
3. Press
to enter the screen.
4. Press
to b
egin the calibration.
3A2040E
17