User's Manual
Operation
Meter Calibra
tion
NOTE: See SetupScreen4for further screen
information, if needed.
When to Calib
rate
• The first time the system is operated.
• Whenever new materials are used in the system,
especially if the materials have viscosities that
differ significantly.
• As part of regular maintenance to retain meter
accuracy.
• Whenever a flow meter is serviced or replaced.
Read Before Calibration
• Meter k-fa
ctor on Setup Screen 4 is updated
automati
cally after the calibration procedure is
completed. You also may manually edit the
k-factor if desired.
• All values on this screen are in cc or cc/pulse,
regardless of the units set in the other Setup
screens.
• Before calibrating the meter, be sure the system
is primed with material.
• Disable alarms before calibration.
Calibration Steps
1. Press
to enter Setup Mode.
2. Press
to move to Setup Screen 4.
3. Press
to enter the screen.
4. Press
to b
egin the calibration.
3A2040E
17