Calibrate Screen 3 initiates a calibration of an
accessory solvent meter. During the test, the Volume
Verification screen will appear.
The meter and
lines must be primed with
solvent befo
re doing the calibration. See
Solvent Mete
r Calibration, page 59 for complete
instructio
ns.
To initiate the calibration, press the Volume Check
button.
The screen displays the volume dispensed. Enter
the amount of solvent dispensed in the Measured
Volume field, or press
to end the test.
After the
Measured Volume is entered, the Accept
Calibra
tion window will appear. Press
to accept
the cal
ibration. Press
to cancel the calibration
and ret
ain the previous K-factor.
Press and hold the Reset button
for 1-2
seconds to reset the volume counter.