Promass Meter Measuring Values
The following are the recommended settings for using the Promass meter with the proportioner. If a value is not men-
tioned, use the default Endress+Hauser setting.
* When changing a value, use or keys to toggle through available selections. Use Enter key to select
and save each value. Only the value flashing can be selected; you may have to select and press for each
digit or unit to move to the next digit/unit.
Promass Meter Function Matrix
The Promass Meter Function Matrix, page 18, shows all the Promass setting values available. The bold values are
the minimum settings required to use the Promass Coriolis Meter with the Graco proportioning system.
Table 2: Promass Meter Values to use with Proportioner
Group Function Value*
System Units Volume Flow cc/m
Unit Volume cc
User Interface Assign Line 1 Volume Flow
Assign Line 2 Totalizer 1 or Operation/System Condition (user’s choice)
Totalizer 1 Assign Totalizer Volume Flow
Reset Total No - default setting
Yes - resets totals, then returns to default
Current Output 1 Assign Current Off
Pulse/Freq. Output Operation Mode Pulse
Assign Pulse Volume Flow
Pulse Value
(K-Factor)
0.020 cc/pulse for Low Flow applications (20-500 cc/min.)
0.061 cc/pulse for Medium Flow applications (500-1500 cc/min.)
0.150 cc/pulse for High Flow applications (1500-3800 cc/min.)
Pulse Width 0.50 ms
Pulse Value 1.00 ms
Output Signal Passive/Positive
Process Parameter Assign LF-Cut Off Volume Flow
On-Val. FL-Cut Off 5.00 cc/min. for Low Flow applications (20-500 cc/min.)
30 cc/min. for Medium & High Flow applications (500-3800 cc/min.)
+ –