User's Manual
Priming and Calibration 
312393G    15 of 50 
Figure 13: Open Dispense Valve (ODV) Screen (C3) 
Key: 
A  Current ODV Adjustment (in mm from the cylinder 
entrance). 
B  Adjust ODV value Option ICON. 
C  Relative ODV Position ICON with respect to 
cylinder entrance (This ICON will move right or 
left at +/- mm transition). 
D  Cylinder Entrance ICON (stationary). 
I  Screen Number (C3) 
J 
 &   keys will navigate to adjacent screens. 
1.  To adjust the ODV position with respect to the 
cylinder entrance, press 
. 
2. Press 
 to increase the value or   to decrease 
the value. Values of – 5.0 mm to + 5.0 mm are 
allowed. 
3. Press 
 to accept the value or   to retain the 
previous value. The default value from the factory 
is – 1.0 mm. 
4. Press 
 twice to return to the Run screen. 
Machine Calibration Shots (C4) 
Read all manufacturer’s warning and material MSDS to know the 
specific hazards of the material used. 
Calibration shots are necessary to program the 
machine to properly dispense the correct amount of 
material. Several short CAL shots (“
”) and several 
long CAL shots (“
”) are required. 
Remove the ratio check nozzle, install the static mixer 
intended to be used, and adjust the air pressure to the 
level intended for use during normal operation. 
The air pressure must be set to operating 
conditions for this step. Significant changes 
in air pressure could adversely affect shot 
accuracy. 
From the Run screen, press 
, then press  , then press 
 3 times. The following screen will be displayed. 
Figure 14: Stroke Calibration (CAL) Screen (C4) 
Key: 
A  Short Shot Column ICON 
B  Last/Current Short Shot weight (mass) entry. 
C  Long Shot Column ICON 
D 
Last/Current Long Shot weight (mass) entry. 
E  Grams (Weight or Mass unit used) 
F  Short Shot ICON 
G  Long Shot ICON 
I  Screen Number (C4) 
J 
 &   keys will navigate to adjacent screens. 
1. Press   to arm the machine to take a short CAL 
shots. The icon will be highlighted (“
”). Press 
 to de-arm the machine. 
2.  Place the empty container below the static mixer, 
and press 
 (green) or the footswitch. 
3. Press 
 and discard the contents of the material 
dispensed 
4.  Repeat the previous 2 steps until the static mixer is 
filled with material. Verify the material dispensed 
from the mixer is mixed properly. 
5.  Place a container on a scale and tare (zero) it. 
6.  Place the tared container below the static mixer, 
and press 
 (green) or the footswitch. 
7. Press 
 to erase existing average weight data. 
8.  Place the container with the dispensed material on 
the tared scale. 
9.  Enter the mass in grams into the HMI, by pressing 
 or   accordingly. 
10. Press 
 to accept the number. The number 
entered will be transferred to the right of the 
average icon (“
”). 
11.  Repeat steps 5 through 10, skipping step 7, several 
times. Each additional shot weight will be averaged 
in the 
 row with the previous shots. 










