Operating Instructions MAVOLUX 5032C/B USB Illuminance Meter 15458 1/03/12
Circular orifice USB Port Light collector Display Key pad Back of display unit: - battery door - serial number - calibration seal for DAkkS- or calibration certificate (if available) Photometer head Back of photometer head: - serial number of display unit - tripod socket ¼ “ Perspective view of MAVOLUX USB 2 GOSSEN Foto- und Lichtmesstechnik GmbH
Contents 1 Page Applications..............................................4 2 2.1 2.2 Display..................................................... 5 Description of Display Elements.............. 5 Key Pad................................................... 5 3 3.1 3.2 Functioning of the MAVOLUX.............. 6 Preparation............................................. 6 Duration of Display – Continuous Operation................................6 4 4.1 4.2 MAVOLUX Operation.............................
1 Applications The MAVOLUX 5032 illuminance meters are handy, easy to use and high precision measuring instruments. They allow the accurate measurement of the illuminance in the unit lx or fc. Combined with the Luminance Attachment, available as optional accessory, the MAVOLUX 5032 can also be used for measuring luminance in the unit cd/m² or fL. The light sensor is color corrected, i.e. its relative spectral responsivity has been adapted to that of the human eye V().
2 Display 2.1 Description of Display Elements 2.
3 Functioning of the MAVOLUX 3.1 Preparation First remove the snap-on cover of the battery compartment at the back of the meter. Insert the battery included in the delivery contents (1.5 V size AA, IEC LR6) in the battery compartment. Care should be taken to place the battery correctly according to the polarity indications “+“ and “-“ in the battery compartment. When the battery warning symbol ( ) appears in the display, the battery must be replaced.
4 MAVOLUX Operation 4.1 Switch On – Make Measurement Press down the on/off key and the MAVOLUX will start up in measuring mode and will immediately be measuring at 2 times per second. The instrument is in the function „AUTO“, i.e. the MAVOLUX will select the best suited measuring range for the existing light level. By pressing one of the range keys, one of the measuring ranges can be set. Scrolling up or down through additional measuring ranges is accomplished by briefly pressing one of the range keys.
4.4 Memory Functions In addition to the display buffer storage, the MAVOLUX has also been provided with a memory space for up to 100 measured values. This function allows the user to make several measurements in the field and then to view them later. The data stored in the memory will be maintained, even if the meter is turned off or if the battery is changed. 4.4.
4.4.2 „Mem“ – Save Measured Values in Memory Hold The measurement measured value which is to be saved will be maintained by the Hold key in the display and may be filed onto the next free memory space by pressing down key Mem. Then, for a few seconds, the display shows the saved measured value alternating with the number of the memory space. After that, the MAVOLUX returns into the measuring mode. If the memory space is filled with all 100 memory spaces, then the display will show “FLL”.
4.4.3 „Mem-Edit“ – Overwrite a Saved Measured Value Hold The MAVOLUX offers the function „Mem Edit“ for correcting a stored value in the memory. Measure again, press the Hold key and the new value is logged in. Then, by pressing one of the range keys, you can scroll through the memory. At the memory space to be corrected press down the key Mem. Then the display shows the saved value for a few seconds alternating with the number of the memory space. Subsequently the MAVOLUX will return to measuring mode.
4.4.4 „Mem Recall“ – Data Memory Recall Mem Start in the function „Measuring“ and press down the key Mem. The last memory space filled will be displayed first. Strike one of the Range keys to scroll through the data memory. The current memory space is displayed as 01-02..., alternating with the measured value saved in the memory.
4.4.5 „Mem-Clear“ – Clear Memory The MAVOLUX must be in function „HOLD“. Pressing down the Range keys simultaneously, will clear the complete measuring data memory. The cancelling is confirmed in the display by three dashes. 5 USB Port – Standard Software The MAVOLUX is equipped with a USB Port which allows the meter to be connected to a PC by the USB cable (in the delivery contents).
6 Accessories 6.1 6.2 Standard Equipment Aluminium Case Battery Protective cover cap Standard Software on CD-ROM USB cable Instruction Manual Optional Accessories Luminance Attachment cd/m² and fL (Order Code M516G) Measures the reflected light, i.e. the luminance of a surface with a measuring angle 1/10 approx. 15°. By a micro switch the MAVOLUX automatically detects the luminance attachment screwed onto the light measuring probe. The measured luminance value is displayed correctly in value and unit.
6.2 Calibration Certificate (Optional) The traceability of the measuring results is guranteed by our calibration reference: Scientific Standard Lamps, type Wi 41G of the PTB (Physikalisch Technische Bundesanstalt Braunschweig – National Standard Institute of Germany). Depending on how the instrument is being used we recommend a recalibration interval between 12 and 24 month. For this purpose please contact our Calibration Service Department (telephone +49 911 8602 172). 6.
8 Specifications Light Sensor Silicon photo diode with V () filter Classification MAVOLUX 5032 USB – Class C according to DIN 5032, Part 7 (CIE 69) MAVOLUX 5032 USB – Class B according to DIN 5032, Part 7 (CIE 69) Measuring Rate 2 measurements per second LCD display Read-out Overload signal 50 mm x 25 mm, (MAVOLUX 5032 B USB: back lighted display) 7 segments, 31/2 digits, 13 mm „OL“ in the display Memory 100 memory space (measured values), display indication „FLL“ = memory full Interface USB 1
Weights and Dimensions Housing Dimensions Weight Light sensor Cable length Plastics Display Unit: 65 x 120 x 19 mm 2.56 x 4.73 x 0.75“ (without carrying case) Photometer Head: 31 x 105 x 30 mm 1.22 x 4.13 x 1.18” tripod socket ¼ “ on bottom side Display unit and photometer head approx. 200 g without battery Light sensitive surface of the diffuser: approx. 10 mm with circular orifice. Location of the reference plane is on the surface of the light collector. Approx. 1.5 m.
Characteristics MAVOLUX 5032 B USB Measuring Quantity Illumination I II III IV V Luminance with luminance attachment for cd/m² or fL I II III IV V Measuring Range in Lux (lx) 0.01... 19.99 0.1... 199.9 1... 1 999 10... 19 900 100... 199 000 in Candela/m² (cd/m²) 0.1... 199.9 1... 1 999 10... 19 990 100... 199 900 1000... 1 999 000 Measuring Range in footcandle (fc) 0.001... 1 999 0.01... 19.99 0.1... 199.9 1... 1 999 10... 19 990 in footlambert (fL) 0.01... 19.99 0.1... 199.9 1... 1 999 10...
V( )-Adaptation (f1´ ) MAVOLUX 5032B USB f1´ < 3 % 18 GOSSEN Foto- und Lichtmesstechnik GmbH
EG – KONFORMITÄTSERKLÄRUNG DECLARATION OF CONFORMITY Dokument-Nr./ Document.No.: Hersteller/ Manufacturer: Anschrift / Address: 102/2004 Produktbezeichnung/ Product name: Typ / Type: Bestell-Nr / Order No: Beleuchtungsstärkemessgerät / Luxmeter / Footcandle meter MAVOLUX 5032 B USB GOSSEN Foto- und Lichtmesstechnik GmbH Thomas-Mann-Str.
Characteristics MAVOLUX 5032 C USB Measuring Quantity Illumination I II III IV Luminance with luminance attachment for cd/m² or fL I II III IV Measuring Range in Lux (lx) 0.1... 199.9 1... 1 999 10... 19 900 100... 199 000 in Candela/m² (cd/m²) 1... 1 999 10... 19 990 100... 199 900 1000... 1 999 000 Measuring Range in footcandle (fc) 0.01... 19.99 0.1... 199.9 1... 1 999 10... 19 990 in footlambert (fL) 0.1... 199.9 1... 1 999 10... 19 990 100... 199 900 Resolution in lx 0.
V( )-Adaptation (f1´ ) MAVOLUX 5032C USB f1´ < 7,5 % GOSSEN Foto- und Lichtmesstechnik GmbH 21
EG - KONFORMITÄTSERKLÄRUNG DECLARATION OF CONFORMITY Dokument-Nr./ Document.No.: Hersteller/ Manufacturer: Anschrift / Address: 101/2004 Produktbezeichnung/ Product name: Typ / Type: Bestell-Nr / Order No: Beleuchtungsstärkemessgerät / Luxmeter / Footcandle meter MAVOLUX 5032 C USB GOSSEN Foto- und Lichtmesstechnik GmbH Thomas-Mann-Str.
GOSSEN Foto- und Lichtmesstechnik GmbH is also a leading provider of further interesting light measuring instruments: - MAVO-MONITOR USB: High quality and precision instrument for measuring luminance, classified according to the Standard Specifications DIN 5032, Part 7, Class B and CIE 69. The measuring sensor to be placed directly on the luminous or back lighted surface may be used for contact measurement such as i.e.
Printed in Germany – Subject to change without notice GOSSEN Foto- und Lichtmesstechnik GmbH Lina-Ammon-Str.22 90471 Nürnberg ● Germany Phone +49 911 8602-181 Fax +49 911 8602-142 E-Mail info@gossen-photo.de www.gossen-photo.