OVSET, OVSET? – Overvoltage Protection Trigger Value
SELECT A and Menu
Function
The triggering threshold reference value required for the OVP
function is set with OVSET.
Setting Command
Syntax: OVSET w
Value range: OVSETmin ≤ w ≤ OVSETmax
Default setting
or after RESET (*RST):
OVSETmax
Query Command
Syntax: OVSET?
Sample response string: OVSET +080,000
Parameters List
Comments
Amongst other causes, overvoltage protection can be triggered
by:
–USET ≥ OVSET (due to manual setting, programming
command, memory recall, sequence run or Uset control signal
to the analog interface)
– Sensing leads with reversed polarity
– Interrupted output lead during sensing mode operation
– Interference from the power consumer
– Parallel connected voltage sources
– Dynamic output voltage overshooting
– Device malfunction or defect
POUT? – Querying Momentary Output Power
SELECT B and Menu
Function
The POUT? function reads out momentary output power as the
product of output voltage and output current.
Query Command
Syntax: POUT?
Sample response string: POUT +XXXXX.X
Measuring range: Due to the fact that the UOUT and IOUT
measuring functions are utilized, the respective measuring ranges
apply for power measurements as well. If one or both of the
measured quantities UOUT and IOUT violate their respective
measuring ranges, the product of POUT (UOUT x IOUT) is
displayed as “–OL” or “OL”, or “+/–999999” is entered to the data
string.
POWER_ON, POWER_ON? –
Response After Power On
Menu
Function
The POWER_ON function determines the status of device
settings after mains power has been switched on.
Setting Command
Syntax: POWER_ON txt
Parameter txt: RST/SBY/RCL/R01 ... R12
Default setting
or after RESET (*RST):
RST
Query Command
Syntax: POWER_ON?
Sample response string: POWER_ON RST
Parameters List
PSET, PSET?
SELECT A
Function
The PSET command is intended for expanded functionality and
has not yet been implemented.
The device’s maximum possible output power can be ascertained
with the PSET? query command.
Depending upon whether the device is operated with 115 or
230 V AC, the response is either 750 W or 1500 W.
Query Command
Syntax: PSET?
Sample response string: PSET +1500.0
Parame-
ters
For:
Device
Type
Setting Range Setting resolution
Nom.
Voltage [V]
OVSETmin.
[V]
OVSETmax.
[V]
Remote
[V]
Manual
[V]
w 60 3.00 80.00 0.02 0.02
Param-
eter
Content Meaning
txt
RST RESET:
Defined default settings are utilized
→ default settings
SBY STANDBY:
Same device settings as before shutdown, but the power output
remains inactive (OUTPUT OFF)
RCL RECALL:
Same device settings as before shutdown – power output remains
in previous state
Rxx Recall a device configuration saved to setup memory under XX