User manual

22 GMC-I Gossen-Metrawatt GmbH
7 Analog Interface
7.1 Connector pin assignments
Interface
Type
Circuit Diagram Func-
tion
Termi-
nal
Meaning
TRG
TRG 1 +
TRG 1 –
TRG 2 +
TRG 2 –
1
2
3
4
Trigger Inputs
Floating digital control inputs for controlling a device function defined by
SEtUP/dPYIF/AnIF/trG 1 txt and SEtUP/dPYIF/AnIF/trG 2 txt
Low signal: –18 V Us + 1 V
High signal: + 4 V Us + 18 V
Current consumption: Is = (Us – 2 V) / 1.47 kΩ
SIG
SIG 1 +
SIG 1 –
SIG 2 +
SIG 2 –
SIG 3 +,
AGND 2
5
6
7
8
9
10
Signal Outputs
Two floating digital status signal outputs
One digital status signal output with reference to AGND (2)
•SIG 1±, SIG 2± and SIG 3+ indicate the statuses defined by
SEtUP/dPYIF/AnIF/SiG 1 txt, SEtUP/dPYIF/AnIF/SiG 2 txt and
SEtUP/dPYIF/AnIF/SiG 3 txt.
Signal type: open collector
Max. switching voltage: 30 V DC
Max. switching current: 20 mA
Interface
Type
Circuit Diagram Func-
tion
Termi-
nal
Meaning
U
H
+15 V
AGND 1
AGND 2
12
12
10
+15 V (output)
This auxiliary voltage output (15 to 18.5 V DC with reference to AGND 1
or AGND 2) can be used to control the trigger inputs, or to supply power
to external components (e.g. reference element for generating control
voltages ).
The output is equipped with electronic current limiting to approximately
60 mA, and is short-circuit proof to AGND 1 and AGND 2.
AGND 1 and AGND 2 (analog ground = reference point)
Reference point for the analog and digital control inputs and outputs
These terminals are internally connected to the minus pole of the power
output via a reversible fuse with a rating of 110 mA.
The following is recommended:
Use AGND 1 as a reference for the analog terminals on the upper
terminal strip (pins 13 through 18).
Use AGND 2 as a reference for the digital terminals on the bottom
terminal strip (pins 1 through 9).S
Q1
1.47 k
Q1
1.47 k
100
100
100
Q2
+T
Q2
Q2
+T
U
I
20 V
60 mA
15 V
+T