Datasheet
GMC-I Messtechnik GmbH 5
SECUTEST BASE / PRO and SECULIFE ST BASE
Test Instruments for
Measuring
Electrical Safety of Devices
Characteristic Values
1
Specified values are only valid for the display at the test instrument. Data transmit-
ted via the USB port may deviate from these values.
2
Known as equivalent leakage current or equivalent patient leakage current from previous standards
3
Protective conductor current, touch current, device leakage current, patient leakage current
4
Protective conductor current, touch current, device leakage current
5
Only with SECUTEST BASE10 (Feature AA02), SECUTEST PRO and SECULIFE ST BASE
6
Only with SECUTEST PRO (Feature I01) and SECULIFE ST BASE
7)
Measurment type IPE clamp and IG clamp
8)
Measurement type IPE AT3 adapter and IG AT3 adapter
9
The measuring range upper limit depends on the selected test voltage.
10)
Due to inrush current limiting components, the voltage at the test socket may be
lower than the measured line voltage.
Key: rdg. = reading (measured value), d = digit(s)
Test Times, Automated Sequence
The test times (parameter „Measurement duration ...“) can be adjusted
in the sequence parameter setting menu for each rotary switch posi-
tion separately. The test times are not tested and calibrated.
Emergency Shutdown During Leakage Current Measurement
As of 10 mA of differential current (can also be set to 30 mA), automatic
shutdown ensues within 100 ms. This shutdown is not effected during
leakage current measurement with clamp or adapter.
Func-
tion
Measured
Quantity
Display Range /
Nominal Range of
Use
Reso-
lution
Nominal
Voltage
U
N
Open-
Circuit
Voltage
U
0
Nom.
Current
I
N
Short-
Circuit
Current
I
K
Inter-
nal
Resis-
tance
R
I
Refer-
ence
Resis-
tance
R
REF
Measuring
Uncertainty
1
Intrinsic Error
1
Overload
Capacity
Value Time
Tests, 62638 (DIN VDE 0701-0702) / IEC 62353 (VDE 0751)
Protective
conductor
resistance
R
PE
1 … 999 mΩ 1mΩ
—
< 24 V
AC or DC
—
>200 mA
AC or DC
> 10 A AC
5)
——
±(15% rdg. + 10 D)
>10D
> 10.0 Ω :
±(10% rdg.+ 10 d)
±(10% rdg.+ 10 d)
>10d
264 V
250 mA
16 A
5)
Cont.
1.00 ... 999 Ω 10 mΩ
10.0 … 30.0 Ω
100 m
Ω
Insulation
resistance
9
RISo
10 … 999 kΩ 1kΩ
50 … 500
VDC
1.0 • U
N
…
1.5 • U
N
>1 mA >2mA — —
±(5% rdg.+ 4 d)
> 10 d
±
(2.5% rdg.+2 d)
> 10 d
264 V Cont.
1.00 … 9.99 MΩ 10 kΩ
10.0 … 99.9 MΩ 100 kΩ
≥ 20 MΩ:
±(10% rdg.+ 8 d)
≥ 20 MΩ:
±
(5% rdg.+4 d)
100 … 300 MΩ 1MΩ
Leakage current,
alternative
measurement
2
IPE, IB, IG, IA
0.0 ... 99 μA1μA
—
50 ...
250 V~
– 20/+10%
—>1.5mA
> 150 k
Ω
1kΩ
±10 Ω
±
(5% rdg.+ 4 d) > 10 d
> 15 mA:
±(10% rdg.+ 8 d)
±
(2% rdg.+2 d) > 10 d
> 15 mA:
±(5% rdg.+ 4 d)
264 V Cont.
100 ... 999 μA1μA
1.00 ... 9.99 mA 10 μA
10.0 ... 30.0 mA 100 μA
Leakage current,
direct
measurement
3
IPE, IB, IG, IA, IP
Only Ip: 0.0 ...
99.9 μA
100 nA
————
1kΩ
±10 Ω
—
±(5% rdg.+ 4 d)
> 10 d
±
(2.5% rdg.+2 d)
> 10 d
264 V Cont.
0.0 ... 99 μA1μA
100 ... 999 μA1μA
1.00 ... 9.99 mA 10 μA
10.0 ... 30.0 mA 100 μA
Leakage current,
differential
current
measurement
4
IPE, IB, IG
0 ... 99 μA1μA
————
1kΩ
±10 Ω
—
±(5% rdg.+ 4 d)
> 10 d
±
(2.5% rdg.+2 d)
> 10 d
264 V Cont.
100 ... 999 μA1μA
1.00 ... 9.99 mA 10 μA
10.0 ... 30.0 mA 100 μA
Function test
Line voltage
U
L–N
10
100.0 … 240.0 V~ 0.1V— — ———— —
±
(2% rdg.+2 d)
264 V Cont.
Load current I
L
0 … 16.00 A
RMS
10mA— — ———— —
±
(2% rdg.+2 d)
16 A Cont.
Active power P
0 … 3700 W 1 W — — — — — — —
±
(5% rdg.+10 d)
>20d
264 V Cont.
20 A 10 min
Apparent power S
0 … 4000 VA 1 VA Calculated value, U
L–N
• I
V
±
(5% rdg.+10 d)
>20d
Power factor PF
with sinusoidal
waveform: cosϕ
0.00 … 1.00 0.01 Calculated value, P / S, display > 10 W ±(10% rdg.+5 d)
Voltage measurement
Probe voltage
(test probe P1 to PE)
, and
0,0 … 99,9 V
100 ... 250 V
100 mV
1 V
————
3 MΩ
——
±
(2 % v.M.+2 D)
300 V
,
and
Cont.
Measurem. voltage
(
sockets
V–COM
6
)
, and
1 MΩ
±
(2 %
rdg.
+2
d
)
> 45 Hz ... 65 Hz
±
(2 %
rdg.
+5
d
)
> 65 Hz ... 10 kHz
±
(5 %
rdg.
+5
d
)
> 10 kHz ... 20 kHz
t
A
PRCD
Time to trip
0.1 ... 999 ms 0.1 ms — — 30 mA — — —
±
5 ms
I
Clamp
Current via
current sensor
WZ12C
[1 mA:1 mV]
(sockets V–COM
6)7
)
1 ... 99 mA
1mA
(1 mV)
— — ———— —
±(2 % rdg.+2 d)
>10D
20 Hz ... 20 kHz
without clamp
253 V Cont.
0,1 ... 0,99 A
0,01 A
(10 mV)
1,0 ... 9,9 A
0,1 A
(100 mV)
10 ... 15 A
1A
(1 V)
I
Leak
Leakage current
via AT3-IIIE
adapter Z745S
6) 8)
0,00 ... 0,99 mA
0,01 mA
— — ———— —
±(2 % rdg.+2 d)
>10D
without adapter
253 V Cont.
1,0 ... 9,9 mA
0,1 mA
10 ... 20 mA
1mA
Temp
Temperature
with Pt100 sensor
– 200,0 ...
+850,0 °C
0,1 °C— <20V– 1,1mA— — — ±(2 % rdg.+1 °C) 10 V Cont.
Temperature with
Pt1000 sensor
– 150,0 ...
+850,0 °C