User manual

72 GMC-I Messtechnik GmbH
Comments
We recommend enabling the following events at the
corresponding registers for an SRQ message:
CME Command Error EXE Execution Error
QYE Query Error OTPA OTP Activated
TSTE Self-Test Error LIME Limit Error
OVPA OVP Activated OCPA OCP Activated
OUTE OUTput Error SEQE Sequence Error
DDTE Define Device Trigger Error
The setting commands required to this end read as follows:
*ESE 52; ERAE 56; ERBE 190; *SRE 52.
As soon as one of these events occurs, the KONSTANTER is
no longer able to execute the desired setting.
RS 232C Serial Interface
Use of the SRQ service request function is not possible via the
RS 232 serial interface.
Similar results can nevertheless be obtained.
Set the PRE register (parallel poll enable) to the same value as
the SRE register (status byte enable register).
Now query and evaluate the IST individual status message.
Query and evaluation can be executed cyclically, and/or each
time data is transmitted to the KONSTANTER .
SSET, SSET? – Switching Function Signal Level
Function and manual operation:
See explanation on page 31.
Programming
Setting command: SSET txt
Parameter: txt (status)
Parameter type: Text (character)
Setting range: OFF signal output n passive high
ON signal output n active low
Example (HP Basic):
OUTPUT 713;"SSET OFF"
Signal output SIG_OUT1/SIG_OUT2 at the
analog interface generates a passive high
signal.
Query
Queries the sequence’s function status.
Programming
Query command: SSET?
Response string: SSET txt
Length: 8 characters, fixed
Parameter: txt (status)
Example (HP Basic):
OUTPUT 713; "SSET?"
ENTER 713; A$
DISP A$
Display: SSET OFF
START_STOP, START_STOP? – Memory Location Start and Stop
Addresses for the SEQUENCE Function
Function and manual operation:
As opposed to manual operation, start and stop memory
addresses are entered to a command.
See explanations on page 32 ff.
Programming
Setting command: START_STOP n1, n 2
Parameter: n1, n2 (start address, stop address)
Parameter type: Integer
Setting range: Start address: 11 to 255
Stop address: 11 to 255
Condition: Start address < stop address
Default setting: after RESET (*RST) : unchanged
Example (HP Basic):
OUTPUT 713;"STA 20, 115"
{sequence start = address 20}
{sequence end = address 115}
Query
Displays selected start and stop memory addresses for the
SEQUENCE function.
Programming
Query command: START_STOP?
Response string: START_STOP n1, n2
Length: 18 characters, fixed
Parameter: n1, n2 (start address, stop address)
Example (HP Basic):
OUTPUT 713; "STA?"
ENTER 713; A$
DISP A$
Display: START_STOP 020,115