ILIM, ILIM? Current Setting Limit Value
Settings
Function and manual operation:
See explanation on page 25.
Programming
Setting command: ILIM v
Parameter: v (value)
Parameter type: Real number
Setting range: 0 v ILIM
max
If the selected value is not a whole number multiple of the step
size, it is rounded off accordingly.
If necessary, the numeric value is rounded off once again for the
4-place digital display.
Default setting: Nominal current (after RESET, *RST)
Example (HP BASIC):
OUTPUT 713; "ILIM 20"
{current setting range limit: 20 A}
Comment
ILIM may not be set to a value of less than ISET!
Any setting command to this effect is not executed.
Bit 1 would otherwise be set in event register B (limit error) and
bit 4 in the standard event register (execution error).
The ILIM function does not effect the adjustment of ISET by
means of a control signal via the analog interface.
Query
Displays the selected final value for current setpoint ISET.
Programming
Setting command: ILIM?
Response string: ILIM v
Length: 13 characters, fixed
Parameter v (value)
Format v: +nn.nnnn
Example (HP Basic):
OUTPUT 713;"ILIM?"
ENTER 713; A$
DISP A$
Display: ILIM +20.0000
IMAX? – Maximum Measured Current Value
Function and manual operation:
See explanation in chapter 4.8.1.
Programming
Query command: IMAX?
Response string: IMAX v
Length: 13 characters, fixed
Parameter: v (value)
Format: +nn.nnnn
Example (HP Basic):
OUTPUT 713; "IMAX?"
ENTER 713; A$
DISP A$
Display: IMAX +02.8550
IMIN? – Minimum Measured Current Value
Function and manual operation:
See explanation in chapter 4.8.1.
Programming
Query command: IMAX?
Response string: IMAX v
Length: 13 characters, fixed
Parameter: v (value)
Format: +nn.nnnn
Example (HP Basic):
OUTPUT 713; "IMIN?"
ENTER 713; A$
DISP A$
Display: IMIN +02.7300
IOUT? – Querying the Momentary Current Value
Function and manual operation:
See explanation on page 23.
Programming
Query command: IOUT?
Response string: IOUT v
Length: 13 characters, fixed
Parameter: v (value)
Format: +nn.nnnn
Example (HP Basic):
OUTPUT 713; "IOUT?"
ENTER 713; A$
DISP A$
Display: IOUT +02.7350
ISET, ISET? – Current Setpoint
Settings
Function and manual operation:
See explanation on page 21.
Programming
Setting command: ISET v
Parameter: v (value)
Parameter type: Real number
Setting range 0 v ILIM
If the programmed value is not a whole
number multiple of the step size, it is
rounded off in accordance with selected
setting resolution.
Device Type
Nom. current [A] 2 3 6 10 12 20
ILIM
max
[A] 2.0 3.0 6.0 10.0 12.0 20.0
Step size [mA] 1.0 1.0 1.0 1.0 1.0 1.0
Device Type
Nom. Current [A] 2 3 6 10 12 20
ISET
max
[A] 2.0 3.0 6.0 10.0 12.0 20.0
Step size [mA] 0.5 1.0 2.0 2.5 3.33 5.0