User manual
GMC-I Messtechnik GmbH 31
tSEt – Memory Location-Specific Dwell Time
Functions
Special dwell time (see also Figure 4.7.3 a):
• For the voltage-current value pair at a specific memory
location within a SEQUENCE
• For all status signals generated for these memory locations
Setting Range
00.00 s The tdEF value is used.
00.01 s to 99.99 s Maximum resolution: 0.01 s
99.99 s Assignment of the specified time to
several consecutive SEQUENCE
memory locations with identical voltage
and current values.
Values include rise time and decay time (see also Figure 4.7.3 a).
Edge times are only assure within a limited load impedance range
(see also Technical Data).
Settings (see sequence settings and execution on page 35)
☞ Procedure for selecting tSEt, (see also chapter 4.7):
! The tSEt display appears in the window for the SEq function,
along with a related numeric parameter [s].
Figure 4.7.3 c Path to tSEt Settings
☞ Set resolution, i.e. the decimal place (cursor position) of the
numeric parameter, by repeatedly pressing the <SELECT> key.
☞ Enter dwell time based on the previously entered resolution
value using the <> key or the <> key.
☞ After selecting the value, jump to the next or the previous
sequence function by pressing the <FUNCTION> key, or by
simultaneously pressing the <FUNCTION> key and the <> key
or the <> key.
☞ The last selected and displayed value for the numeric dwell
time parameter is used as a default setting.
SSEt – Setting a Digital Switching Function
Functions
• Preparation of the SSP KONSTANTER for the read-out of a
status signal from a specific memory location.
• The duration of the signal depends upon the dwell time, tSEt
or tdEF, selected for the memory location.
• The signal is read out to the analog interface’s SiG1 and/or SiG2
outputs.
• Prerequisite: The desired output (SiG1, SiG2 or SiG1 and SiG2)
must be prepared with the SSEt parameter in the AniF function
menu (see also page 29).
Setting Parameter
☞
OFF: No status signal
☞ ON: Signal outputs prepared for read-out of a status signal
Settings (see sequence settings and execution on page 35)
☞ Procedure for selecting SSEt, (see also chapter 4.7 “Settings”):
! The SSet display and a related text parameter appear in the
SEq function window.
Figure 4.7.3 d Path to Selection of the SSEt Text Parameter
☞ Select the desired parameter with the <> key or the <> key.
☞ Acknowledge your selection with the <ENTER> key.
! Parameters can be scrolled through in a loop with the <> key
or the > key.
☞ Advance through the SEq functions menu by pressing the
<FUNCTION> key or the <FUNCTION><> and <> keys, or
press the <CE/LOCAL> key to jump back to the function group
menu (FUNCTION).
Select the text parameter
from an options loop
ENTER
CE/LOCAL FUNCTION
SELECT
Cursor position
Resolution of
numeric parameter
*
From the SEq function group
Select the text parameter
from an options loop
From the SEq function group
*
ENTER
CE/LOCAL FUNCTION