SiG1 – Signal Output 1
Functions
Digital open collector output with reference to AGND, max.
switching voltage min. 30 V DC, max. switching current min.
20 mA (e.g. ext. load shedding relay).
Indicates the status of the power output.
If an event occurs which is associated with the selected
parameter, a corresponding signal is generated at the output.
Used to link the OUTPUT function via the trigger input in the
case of parallel or series (master-slave) connection of several
SSP KONSTANTERs.
Control of an external output relay is possible.
As a status signal for monitoring devices
As a freely assignable switching output
1
The SSEt parameter must be set to ON for a certain memory location in the SEQUENCE function
group.
When the SEQUENCE is run, an signal is additionally generated
for the read-out of Uset and Iset when this particular memory
location is executed.
The signal is available at the outputs of the analog interface
(SiG1 / SiG2 / SiG1 + SiG2) for the respectively selected dwell
time tSEt or tdEF.
Setting
Procedure for selecting Sig1: see also
chapter 4.7, “Settings”.
! The Sig1 display and a related text
parameter (status) appear in the
AnIF function window.
Figure 4.7.2 b Path to Selection of the SiG1 Text Parameter
SiG2 – Signal Output 2
Function and Parameters
Identical to SiG1, page 29
Settings
Identical to SiG1, page 29
Setting Parameters SIG1 OUT Status
OFF
Always passive
high
ON
Always active low
OUT
Power output:
OUTPUT ON
OUTPUT OFF
Passive high
Active low
ModE
Control mode signal:
Constant current regulation
Overload (Pmax)
Constant voltage, output Off
Active low
Active low
Passive high
SEq
Status with running sequence
Active low
SSEt
If SSEt is set to ON for special
memory location in the SEQUENCE
1
,
otherwise
Active low
Passive high
U_LO
Umeas < Uset (reference value memory
Umeas Uset254)
Active low
Passive high
U_HI
Umeas > Uset (reference value memory
Umeas Uset255)
Active low
Passive high
I_LO
Imeas < Iset (reference value memory
Imeas Iset254)
Active low
passive high
I_HI
Imeas > Iset (reference value memory
Imeas Iset255)
Active low
Passive high
Select the text parameter
from an options loop
From the AniF function group
*
ENTER
CE/LOCAL FUNCTION