User manual
GMC-I Messtechnik GmbH 29
SiG1 – Signal Output 1
Functions
• Digital open collector output with reference to AGND, max.
switching voltage min. 30 V DC, max. switching current min.
20 mA (e.g. ext. load shedding relay).
Indicates the status of the power output.
• If an event occurs which is associated with the selected
parameter, a corresponding signal is generated at the output.
• Used to link the OUTPUT function via the trigger input in the
case of parallel or series (master-slave) connection of several
SSP KONSTANTERs.
• Control of an external output relay is possible.
• As a status signal for monitoring devices
• As a freely assignable switching output
1
The SSEt parameter must be set to ON for a certain memory location in the SEQUENCE function
group.
☞ When the SEQUENCE is run, an signal is additionally generated
for the read-out of Uset and Iset when this particular memory
location is executed.
☞ The signal is available at the outputs of the analog interface
(SiG1 / SiG2 / SiG1 + SiG2) for the respectively selected dwell
time tSEt or tdEF.
Setting
☞
Procedure for selecting Sig1: see also
chapter 4.7, “Settings”.
! The Sig1 display and a related text
parameter (status) appear in the
AnIF function window.
Figure 4.7.2 b Path to Selection of the SiG1 Text Parameter
SiG2 – Signal Output 2
Function and Parameters
☞
Identical to SiG1, page 29
Settings
☞
Identical to SiG1, page 29
Setting Parameters SIG1 OUT Status
☞ OFF
Always passive
high
☞ ON
Always active low
☞ OUT
Power output:
OUTPUT ON
OUTPUT OFF
Passive high
Active low
☞ ModE
Control mode signal:
Constant current regulation
Overload (Pmax)
Constant voltage, output Off
Active low
Active low
Passive high
☞ SEq
Status with running sequence
Active low
☞ SSEt
If SSEt is set to ON for special
memory location in the SEQUENCE
1
,
otherwise
Active low
Passive high
☞ U_LO
Umeas < Uset (reference value memory
Umeas Uset254)
Active low
Passive high
☞ U_HI
Umeas > Uset (reference value memory
Umeas Uset255)
Active low
Passive high
☞ I_LO
Imeas < Iset (reference value memory
Imeas Iset254)
Active low
passive high
☞ I_HI
Imeas > Iset (reference value memory
Imeas Iset255)
Active low
Passive high
Select the text parameter
from an options loop
From the AniF function group
*
ENTER
CE/LOCAL FUNCTION