Operation Manual
8 GMC-I Messtechnik GmbH
1.4 List of Possible Options and Standard Types
1)
Each measured value is documented in this case, as opposed to the results of a
test sequence for which only the poorest value for each given test is displayed.
(via the PSI module, the SECUSTORE storage adapter or a PC)
2)
Adapter set for international use (comes with Feature B01)
3)
only possible with Feature J01
4)
if adapter (feature B11) is applied: HV-DC max. 1.5 kV DC
5)
without function test values and without comments on DUT
6)
Testing of patient leakage currents and patient auxiliary currents is not possible
without Feature J01
Standard Types
Features which are additionally required can be retrofitted by our service department upon request, address see chapter 23.
Features 00 01 02 03 04 05 06 07 08 09 10 11 XX
Mains Connection for
Respective Country
B D
D + ser-
vice socket
UK
4)
F/CZE DK
4)
China/AUS
4)
CH
adapter
set
2)
4)
User Interface Language C DUKF I ECZENL
High Voltage Testing
HV-DC F without
max.
6.126 kV DC
( 4KVAC)
AC Test Current 50/60 Hz
for Protective Conductor
Measurement
G 10 A 25 A without
SECULIFE ST: 10 + 2 Plugs
for Application Parts
J without
6)
with
Test Sequence
for IEC 60601
KA without with
3)
Data Memory
for up to 125 Tests
5)
KB without with
Recognition of
Probe to Protective Conductor
KD without with
Direct Printing after each
Measurement for Automatic
Test Sequences
1)
Read-Out via RS232
KE without with
Type Designation / Feature Combination Article number
SECUTEST SIII+H including selectable test current 200 mA DC or 25 A AC (G01), including high voltage testing up to 6 kV DC (F02)
including sequences for IEC 61010, IEC 60335, IEC 60950, including data memory for up to 125 tests
5)
(KB01) M7010-V013