Operation Manual
GMC-I Messtechnik GmbH 63
Mechanical Design
Display multiple dot matrix display, 128 x 128 pixels
Dimensions test instruments without high-voltage module:
LxWxH: 292 mm x 138 mm x 243 mm
test instruments with high-voltage module:
LxWxH: 292 mm x 138 mm x 300 mm
Weight standard test instrument: approx. 4.5 kg
instrument with HV test: approx. 5.24 kg
instrument with 25 A PE test: approx. 5.5 kg
instr. with 25 A PE & HV test: approx. 5.9 kg
Protection housing: IP 40
terminals: IP 20 per
DIN VDE 0470, part 1/EN 60529
Extract from table on the meaning of IP codes
Service socket (20) – Terminal data
(Prerequisite: Feature B01)
Line Voltage 103.5 V ... 126.5 V or 207 V ... 253 V
Line Frequency 50 Hz or 60 Hz
Current output protected by building installation
(16 A; fed through to the mains plug)
High-Voltage Test
(Prerequisite: Feature F02 or SECUTEST SI II+H)
Transducer
Test Duration as long as START key is pressed (max. 60 sec.)
Measuring
Maximum Test Voltage
SC I* DUTs 1.5 kV
SC II DUTs 4 kV
* devices with protective conductor terminal
IP XY
(1
st
digit X)
Protection against
foreign object entry
IP XY
(2
nd
digit Y)
Protection against the
penetration of water
0 not protected 0 not protected
1 50.0 mm
1 vertically falling drops
2 12.5 mm
2
vertically falling drops with
enclosure tilted 15
3 2.5 mm
3spraying water
4 1.0 mm
4 splashing water
Nominal Voltage, AC
Open-Circuit Voltage, DC
U
N~
adjustable in 10 V steps
in 100 V steps
Uo
0.5 0.99 kV
1 4.0 kV
((U
N~
· 1.5) · 1.011) + 60 V
Intrinsic Error, Uo Uo 1.5%
Nominal Current per DIN VDE 0104 < 3.5 mA DC
Short-Circuit Current discharge current > 5 A at 6 kV
Resistance to Interference
Voltage
none
Measuring Range Display Range Intrinsic Error, Uo
0 Uomax 0.000 > 10.00 kV DC 1.5% rdg. + 2 digits