Operation Manual
54 GMC-I Messtechnik GmbH
15.12 Testing in Accordance with IEC 62353/VDE 0751
The following measurements can be performed in accordance with the
above mentioned standard:
• Protective conductor measurement R
PE
Test current: 200 mA DC,
Test current: 10 A AC
(Feature G00
)
Test current: 25 A AC
(Feature G01 or
SECUTEST SIII+H
)
• Insulation measurement (can be additionally activated)
– R-INS LN-PE
(insulation resistance, LN to protective conductor)
– R-INS AP-PE (insulation resistance, app. part to protective conductor)
• Equivalent device leakage current I
EDL
• Equivalent patient leakage current I
EPL
• Device leakage current (direct or residual current)
• Contact current
• Patient leakage current
(direct or mains at application part ) (only with Feature J01)
Leakage current is converted to reference voltage (see limit values chapter
8 on page 18). Reference voltage must be adapted to the supply voltage
range.
Check connection parameters and start test.
To
Socket This is the default setting. Refer to chapter 7 on page 17 for
other types of connection.
Class If the DUT has been connected to the test socket, safety
class recognition is performed (SC I or SC II).
In all other cases, or if it is not clear whether or not all
exposed, conductive parts are connected to one another or
to the protective conductor, the safety class can be selected
manually.
Type Select the device type (DUT) from the list.
If “old devices” is selected, limit values from DIN VDE 0701-
0702 are used.
AP-Type:... (BF): Application parts are recognized automatically, and can
be manually changed as well:
Select the App. prt. ... line with the
or key, acknowl-
edge with the
key and change with the or key.
02: The number of configured groups is displayed here.
After selecting the App. prt. ... line with the or key,
acknowledging twice with the key and then pressing the
or key, the “Configure application parts” menu appears
(see chapter 15.13 on page 56).
ID No. See parameters database in chapter 15.2 on page 37.
Setup Refer to chapter 15.2 on page 37 regarding setup of the
measuring sequence.