15.11 Testing Devices in Accordance with EN 60335
The following tests can be performed in accordance with VDE 0700, part
500, (as part of DIN EN 50106:1998) in compliance with special rules for
routine testing of devices for which EN 60335-1 and EN 60967 apply:
• Testing of the protective conductor connection by means of resistance
measurement R
PE
(permanent connection or with plug)
Test current: 10 A AC
(Feature G00
)
Test current: 25 A AC
(Feature G01 or
SECUTEST SIII+H
)
•
Testing for dielectric strength by applying DC high-voltage (1.5-fold value)
(Prerequisite: Feature F02 or SECUTEST SIII+H)
•Function test
The following tests can be performed in accordance with
EN 60335-1:1994:
• Testing for dielectric strength by applying high-voltage
(Prerequisite: Feature F02 or SECUTEST SIII+H)
• Testing for Equivalent Leakage Current
Additional testing options:
• Insulation resistance measurement R
INS
• Residual current
Check connection parameters and start test.
If desired, the test sequence can be adapted to the respective device
under test in the following menu in Setup... under Sequence....
To Socket This is the default setting. Refer to chapter 7 on page
17 for other types of connection.
Class If the DUT has been connected to the test socket,
safety class recognition is performed (SC I or SC II).
Otherwise, the safety class must be specified manu-
ally.
Type A specific DUT type can be selected from a list of
devices under test, if these have been entered in the
setup menu.
ID No. An individual serial number (up to 10 characters) can
be entered with the keypad at the (P)SI module
(accessory), or can be read in with a barcode scan-
ner (accessory).
If an incorrect entry is made: Only complete line can
be deleted, and deletion is only possible with the
key at the instrument.
Setup Refer to chapter 15.2 on page 37 regarding setup of
the measuring sequence.