Operation Manual
50 GMC-I Messtechnik GmbH
15.10 Testing Devices in Accordance with EN 61010
The following measurements can be performed in accordance with the
above mentioned standard. Connection must be made via the test
socket:
• Protective conductor R
PE
Test current: 10 A AC
(Feature G00
)
Test current: 25 A AC
(Feature G01 or
SECUTEST SIII+H
)
• Testing of dielectric strength by applying high voltage with direct volt-
age (1.5-fold value)
(Prerequisite: Feature F02 or SECUTEST SIII+H)
• Insulation measurement R
INS
(can be deactivated)
• Housing leakage current under normal condition I
HLNC
and single-fault
condition I
HLSF
with interrupted protective conductor
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page
17 for other types of connection.
Class If the DUT has been connected to the test socket,
safety class recognition is performed (SC I or SC II).
Otherwise, the safety class must be specified manu-
ally.
Type A specific DUT type can be selected from a list of
devices under test, if these have been entered in the
setup menu.
ID No. An individual serial number (up to 10 characters) can
be entered with the keypad at the (P)SI module
(accessory), or can be read in with a barcode scan-
ner (accessory).
If an incorrect entry is made: Only complete line can
be deleted, and deletion is only possible with the
key at the instrument.
Setup Refer to chapter 15.2 on page 37 regarding setup of
the measuring sequence.