15.6 Testing Devices in Accordance with DIN VDE 0701-0702:2008
The following measurements can be performed in accordance with the
above mentioned standard:
• Protective conductor R
PE
(permanent connection or with plug)
Test current: DC 200 mA
Test current: 10 A AC
(Feature G00
)
Test current: 25 A AC
(Feature G01 or
SECUTEST SIII+H
)
• Insulation measurement R
INS
(can be deactivated, e.g. if danger
exists that voltage sensitive components in data processing systems
might be damaged) plus equivalent leakage current
or
• Contact current for safety class II
or
• Residual current
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page
17 for other types of connection.
Class If the DUT has been connected to the test socket,
safety class recognition is performed (SC I or SC II).
Otherwise, the safety class must be specified manu-
ally.
Ext. Cable
WITH EL1
x: Extension cables or connector cables which are
longer than 5 m can be tested with the help of the
EL1 adapter (optional), either separately or in combi-
nation with a device, see chapter 15.7.
R-INS LN-SL x: Insulation resistance measurement is performed.
ID No. See parameters database in chapter 15.2 on page
37.
Setup See chapter 15.2 on page 37 regarding setup of the
measuring sequence.