15.6 Testing Devices in Accordance with DIN VDE 0701-0702:2008
The following measurements can be performed in accordance with the
above mentioned standard:
Protective conductor R
PE
(permanent connection or with plug)
Test current: DC 200 mA
Test current: 10 A AC
(Feature G00
)
Test current: 25 A AC
(Feature G01 or
SECUTEST SIII+H
)
Insulation measurement R
INS
(can be deactivated, e.g. if danger
exists that voltage sensitive components in data processing systems
might be damaged) plus equivalent leakage current
or
Contact current for safety class II
or
Residual current
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page
17 for other types of connection.
Class If the DUT has been connected to the test socket,
safety class recognition is performed (SC I or SC II).
Otherwise, the safety class must be specified manu-
ally.
Ext. Cable
WITH EL1
x: Extension cables or connector cables which are
longer than 5 m can be tested with the help of the
EL1 adapter (optional), either separately or in combi-
nation with a device, see chapter 15.7.
R-INS LN-SL x: Insulation resistance measurement is performed.
ID No. See parameters database in chapter 15.2 on page
37.
Setup See chapter 15.2 on page 37 regarding setup of the
measuring sequence.