15.4 Testing Devices in Accordance with DIN VDE 0701, Part 1
You can either change over to testing in accordance with the new
VDE 0701-0702 standard or perform the following measurements in ac-
cordance with the above-mentioned standard:
• Protective conductor measurement R
PE
(permanent connection or with plug)
– Part 1: Test current: 200 mA DC
– Appendices:
Test current: 10 A AC (Feature G00
)
Test current
: 25 A AC (Feature G01 or
SECUTEST SI II+H
)
• High-voltage test as supplement (Appendix E)
(Prerequisite: Feature F02 or SECUTEST SIII+H)
• Insulation resistance measurement R
INS
Part 1
The following safety class I through III appliances and electric equipment
can be tested with this standard, for example:
• Devices with electric motors
• Electric heating devices
• Electric tools
• Light fixtures
• Stereo equipment, television sets
SC I extension cables can also be tested (see chapter 15.7 on page 46).
Appendices (previously part 260)
Appendix E: Electric tools
Check connection parameters and start test.
To Socket This is the default setting. Refer to chapter 7 on page
17 for other types of connection.
Class If the DUT has been connected to the test socket,
safety class recognition is performed (SC I or SC II).
Otherwise, the safety class must be specified manu-
ally.
Ext. Cable
WITH EL1
x: Extension cables or connector cables which are
longer than 5 m can be tested with the help of the
EL1 adapter (optional), either separately or in
combination with a device, see chapter 15.7.
ID No.
See parameters database in chapter 15.2 on page 37.
Setup See chapter 15.2 on page 37 regarding setup of the
measuring sequence.
Note
+
Extension cables can only be tested with the VDE 0701-0702
(VDE 0701 part 1) parameter settings, if the EL1 accessory
adapter is used (see chapter 15.7 on page 46).