All of the possible sequence settings for all of the regulations are listed
below.
Select the Setup... menu from the initial program window and
acknowledge.
Sequence 1 (repair tests, periodic testing)
store All of the settings in the setup menu, i.e. the configu-
ration of the measuring parameters and the current
limit values can be saved with this command for the
selected test standard. These values remain active
even after setting the selector switch to a different
position, and after disconnection from mains power.
Sequence See page 39.
Limit Values See chapter 4.6 on page 14.
Database Start with ID no.
x: Before each measurement is started, an entry
prompt appears requesting entry of an ID no.
An individual number can be entered (max. 20
characters) with the keypad at the (P)SI module
(optional), read in with a barcode scanner
(optional) or directly selected from a list.
If an incorrect entry is made:
Only complete lines can be deleted, and deletion
is only possible with the key at the instrument.
ID no. = test sequence (Feature KB01 or
SECUTEST
SIII+H
).
See chapter 16 on page 58.
Additional Parameters
R-PE with clip x: Protective conductor resistance can be
determined with the help of the WZ12C current clip.
High-voltage x: A high-voltage test is conducted
(prerequisite: Feature F02 or SECUTEST SIII +H)
Sequence 2 (type tests)
Type: ........ Sequences 1 and 2 differ primarily as regards the
type parameter. The user can set up any desired type
of device under test with sequence 2, for which the
same limit values and measuring parameters always
apply. Up to 125 different types can be defined. DUT
types are predefined for sequence 1 (EN 60601).
Designations made up of alphanumeric characters
(max. 10 characters) can be entered at the keypad of
a (P)SI module, or at a PC with a terminal program.
The test sequence, including all limit values, is saved
together with the selected type. We recommend
using the characters which identify the respective
standard as the initial characters in the designation. If
a type is queried from the initial program window
which does not belong to the currently selected stan-
dard, these characters identify the test sequence as
belonging to another standard.
Limit Values See chapter 4.6 on page 14.
Sequence See page 39.
Store as
Default All of the settings in the setup menu, i.e. the configu-
ration of the measuring parameters and the current
limit values can be saved with this command.
These values remain active even after setting the
selector switch to a different position, and after dis-
connection from mains power.
Delete The type which has been selected in the initial pro-
gram window can be deleted.